FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2141912
·
Received June 28, 2011
Report
- Report Number
- 2050012-2011-02651
- Event Type
- Malfunction
- Date Received
- June 28, 2011
- Date of Event
- May 28, 2011
- Report Date
- May 28, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A BCE FIELD SERVICE ENGINEER (FSE) FOUND THE FLUID LEAK COMING FROM THE SAMPLE WASH COLLAR. FSE CLEANED THE WASH COLLAR WASTE VALVE AND CLEANED ALL CUVETTES. REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) CUSTOMER TECHNICAL SUPPORT (CTS) IN REGARDS TO FAILING WATER BLANK TESTS, CARTRIDGE CHEMISTRY LEAK, AND NO VACUUM AT THE WASH COLLAR. NO INJURY OR EXPOSURE WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC SYNCHRON® CHEMISTRY ANALYZER | DISCRETE CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | N/A | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |