FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2141912 · Received June 28, 2011

Report

Report Number
2050012-2011-02651
Event Type
Malfunction
Date Received
June 28, 2011
Date of Event
May 28, 2011
Report Date
May 28, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A BCE FIELD SERVICE ENGINEER (FSE) FOUND THE FLUID LEAK COMING FROM THE SAMPLE WASH COLLAR. FSE CLEANED THE WASH COLLAR WASTE VALVE AND CLEANED ALL CUVETTES. REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC (BEC) CUSTOMER TECHNICAL SUPPORT (CTS) IN REGARDS TO FAILING WATER BLANK TESTS, CARTRIDGE CHEMISTRY LEAK, AND NO VACUUM AT THE WASH COLLAR. NO INJURY OR EXPOSURE WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC SYNCHRON® CHEMISTRY ANALYZER DISCRETE CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. N/A N/A

Patients

Seq Age Sex Outcome Treatment
1