COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-09041
- Event Type
- Malfunction
- Date Received
- June 28, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 15, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY AWAITING EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DISPLAY MALFUNCTION WAS CONFIRMED BY SERVICE. THE CAUSE OF THE REPORTED CONDITION WAS DETERMINED TO BE A FAULTY MAIN DISPLAY. THE MAIN DISPLAY WAS REPLACED TO FIX THE REPORTED CONDITION. ADDITIONAL INFORMATION: THIS DEVICE IS AN UNREMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE HAS BEEN ESCALATED TO CAPA. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A BAD SCREEN. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED IN BIOMED SERVICE. ACCORDING TO THE HOSPITAL REPRESENTATIVE, NO PATIENT INVOLVEMENT, INJURY, OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS PUMP IS CURRENTLY UNKNOWN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |