FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2141874 · Received June 28, 2011

Report

Report Number
2531779-2011-04512
Event Type
Malfunction
Date Received
June 28, 2011
Report Date
May 31, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP WAS RETURNED TO ANIMAS AND EVALUATED. A LEAK TEST WAS PERFORMED AND THE DISPLAY LENS WAS FOUND TO HAVE LEAK. THE KEYPAD COVER WAS LOOSE ALONG THE SIDE OF KEYPAD BESIDE THE UP AND DOWN ARROW BUTTONS. TESTING CONFIRMED INTERMITTENT RESPONSES TO BUTTON PRESSES ON THE KEYPAD FOR ALL BUTTONS. MULTIPLE BUTTON PRESSES WERE REQUIRED BEFORE THE BUTTONS WOULD ENGAGE. THE BUTTONS DID NOT HAVE NORMAL SPRING BACK AND CLICK. THE KEYPAD WAS REMOVED AND ADHESIVE WAS FOUND UNDER THE BUTTON CONTACTS.

Description of Event or Problem · 1

ON (B)(6), 2011, THE LAY-USER/PATIENT CONTACTED ANIMAS ON BEHALF OF HER DAUGHTER (THE PATIENT) ALLEGING THAT KEYPAD BUTTON PRESSES DID NOT ACTIVATE DESIRED PUMP FUNCTIONS. THE PATIENT CLAIMED THAT THE PUMP WAS NOT RESPONDING TO BUTTON PRESSES. THE PATIENT DID NOT ALLEGE ANY HARM OR INJURY BECAUSE OF THE REPORTED ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 16 YR