FDA Adverse Event Malfunction Summary report: N

ROTATING MULTIPLE CLIP APPLIER

MDR report key: 2141827 · Received June 27, 2011

Report

Report Number
3005075853-2011-02608
Event Type
Malfunction
Date Received
June 27, 2011
Date of Event
May 30, 2011
Report Date
June 3, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL INFORMATION: AFTER SEVERAL REQUESTS, THE DEVICE WAS NOT RECEIVED FOR ANALYSIS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A COLON (COELIOSCOPY) PROCEDURE, THE CLIPS WERE DELIVERED ACCROSSED. THE CLIPS WERE PLACED ACCROSSED INTO THE JAWS PRIOR TO FIRING. THE INCIDENT OCCURRED FROM THE FIRST CLIP. AT EACH ACTIVATION (MESENTERIC VEIN), THE CLIP WOULDN'T CLOSE AND FELL INTO THE PATIENT. THERE WAS NO TORQUING OR TWISTING OF THE DEVICE, NO UNEXPECTED NOISE HEARD OR RESISTANCE FELT WHILE FIRING THE TRIGGER. THE DEVICE WAS NOT FIRED AFTER THIS INCIDENT. AFTER THE FOURTH CLIP, THE SURGEON TOOK OUT THE CLIPS WHICH HAD FALLEN INTO THE PATIENT. THEN HE USED A SECOND LIKE DEVICE TO PERFORM THE PROCEDURE, WHICH ENDED WITHOUT FURTHER ISSUE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ROTATING MULTIPLE CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. NA G4UF8A

Patients

Seq Age Sex Outcome Treatment
1