FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2141813 · Received June 27, 2011

Report

Report Number
6000001-2011-08994
Event Type
Malfunction
Date Received
June 27, 2011
Date of Event
June 6, 2011
Report Date
June 7, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
PMA / PMN Number
K063696
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY AWAITING EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 543 WAS DISCOVERED AND CONFIRMED IN THE PUMP'S EVENT HISTORY BY A BAXTER SERVICE TECHNICIAN. THE ROOT CAUSE OF THIS CONDITION WAS ASSIGNED TO A FAULTY USER INTERFACE MODULE PRINTED CIRCUIT BOARD AND DEPLETED MAIN BATTERIES. THE MAIN BATTERIES, BATTERY HARNESS, AND USER INTERFACE MODULE PRINTED CIRCUIT BOARD WERE REPLACED TO CORRECT THIS CONDITION. THIS IS INVOLVING A PUMP WITH SOFTWARE VERSION 6.13.90 WHICH IS CATEGORIZED AS A COLLEAGUE 2006. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION.

Description of Event or Problem · 1

THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH FAILURE CODE 543. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS CONDITION OCCURRED. ACCORDING TO THE HOSPITAL REPRESENTATIVE, THERE WAS NO PATIENT INVOLVEMENT. NO PATIENT INJURY OR MEDICAL INTERVENTION HAD BEEN REPORTED RELATED TO THE DEVICE. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION OF THIS DEVICE IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1