FDA Adverse Event Malfunction Summary report: N

COLLEAGUE PRE P1.7

MDR report key: 2141795 · Received June 27, 2011

Report

Report Number
6000001-2011-08989
Event Type
Malfunction
Date Received
June 27, 2011
Date of Event
June 3, 2011
Report Date
June 3, 2011
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT CONTAIN A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US. A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Additional Manufacturer Narrative · 1

(B)(4). THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A FAILURE CODE 804:26 WAS CONFIRMED DURING AN EVENT HISTORY REVIEW. THIS CONDITION WAS DUE TO A DEFECTIVE PUMP HEAD MODULE (PHM) ON CHANNEL A. THE PUMPHEAD MODULE WAS REPLACED TO FIX THE REPORTED CONDITION. THIS ISSUE HAS BEEN ESCALATED TO CAPA. THIS INVOLVED A COLLEAGUE INFUSION PUMP WITH A USER INTERFACE MODULE MASTER SOFTWARE VERSION 5.08.92.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT A COLLEAGUE INFUSION PUMP WHICH EXPERIENCED FAILURE CODE 808:02. THE PATIENT INVOLVEMENT IS UNKNOWN. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE. THE USER INTERFACE MODULE MASTER SOFTWARE VERSION IS CURRENTLY UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 COLLEAGUE PRE P1.7 PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1