FLOGARD

MDR report key: 2141793 · Received June 27, 2011

Report

Report Number
6000001-2011-08988
Event Type
Malfunction
Date Received
June 27, 2011
Date of Event
May 31, 2011
Report Date
June 3, 2011
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K915523
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A FLO-GARD INFUSION PUMP A FALSE AIR IN LINE ALARM ON THE LEFT SIDE WAS CONFIRMED AND DUPLICATED BY BAXTER SERVICE PERSONNEL. THE ROOT CAUSE OF THIS CONDITION WAS DETERMINED TO BE DAMAGE TO THE AIR SENSOR PRINTED CIRCUIT BOARD ON PUMP 1. THE AIR SENSOR ON PUMP 1 WAS REPLACED TO CORRECT THIS CONDITION. A SERVICE HISTORY REVIEW REVEALED NO PREVIOUS SERVICE EVENTS WERE RELATED TO THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). A FOLLOW-UP MEDWATCH REPORT WILL BE SUBMITTED WHEN THE EVALUATION RESULTS OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE CONTACTED BAXTER TO REPORT A FLOGARD PUMP WHICH EXPERIENCED A FALSE AIR IN LINE ALARM. THE EVENT OCCURRED IN THE GENERAL PATIENT WARD. THERE WAS NO ADVERSE EVENT, PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS REPORT. THERE IS NO FURTHER COMPLAINT INFORMATION AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE CORPORATION

Patients

Seq Age Sex Outcome Treatment
1