VERTEX RECONSTRUCTION SYSTEM
Report
- Report Number
- 1030489-2011-00821
- Event Type
- Injury
- Date Received
- June 27, 2011
- Date of Event
- May 31, 2011
- Report Date
- November 2, 2011
- Manufacturer
- WARSAW ORTHOPEDICS, INC.
- Product Code
- KWP
- PMA / PMN Number
- K052376
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IT
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4). NEITHER THE DEVICE NOR FILMS OF APPLICABLE IMAGING STUDIES WERE RETURNED TO THE MANUFACTURER FOR EVALUATION. THEREFORE, WE ARE UNABLE TO DETERMINE THE DEFINITIVE CAUSE OF THE REPORTED EVENT. A REVIEW OF THE DEVICE HISTORY RECORDS FOR THIS DEVICE DID NOT REVEAL ANY NON-CONFORMANCES TO SPECIFICATION OR DEVIATIONS IN PROCEDURE WHICH MIGHT CONTRIBUTE TO THE REPORTED EVENT.
ANALYSIS OF THE RETURNED DEVICE SHOWS THAT THE CROWN IS SHOWING IMPACTIONS DUE TO TIGHTENING WITH A SPINAL ROD. THE DEFORMATIONS ARE ASYMMETRICAL FOR THREE MAS REFERENCE (B)(4), CONSISTENT WITH THE TRANSMISSION OF FLEXURAL LOADS THROUGH THE CONNECTION WHEREAS THE DEFORMATIONS ARE SYMMETRICAL FOR TWO MAS REFERENCE (B)(4), CONSISTENT WITH THE PROPER TIGHTENING OF THE CONNECTION. ALL SETSCREWS (FOR DOMINOS AND FOR MAS) PRESENT WORN SURFACES AT THE FLAT BOTTOM WHICH ARE CONSISTENT WITH THE TIGHTENING OF THE RODS. THE SHORTEST ROD PRESENTS ONE MARK COMING FROM THE MAS SETSCREWS AND ONE MARK COMING FROM THE DOMINO SETSCREW AT ITS POSTERIOR SIDE. AT ITS OPPOSITE ANTERIOR SIDE, THIS ROD PRESENTS ONE CONTACT WITH THE MAS CROWN AND TWO LINES OF CONTACT WITH THE DOMINO ROD CANAL. THE ROD END CLOSED TO THE CONTACT WITH THE DOMINO HAS BEEN CUT WITH A ROD CUTTER. THE ROD IS BROKEN AT THE OPPOSITE END (CLOSED TO THE MAS CONTACT). THE FRACTURE APPEARANCE IS TYPICAL OF FATIGUE DAMAGING OF THE MATERIAL WITH VISIBLE LINES OF REST ACROSS THE SECTION. THE LONGEST ROD IS BROKEN IN TWO AT THE JUNCTION BETWEEN THE Ø3.2 AND Ø3.6 PORTIONS. THE FRACTURE IS ORTHOGONAL TO THE ROD DIRECTION. BOTH ENDS OF THE LONGEST ROD HAVE BEEN CUT WITH A ROD CUTTER. THE Ø3.6 PORTION PRESENTS ONE MARK COMING FROM A SETSCREW AT ITS POSTERIOR SIDE. THE BROKEN SURFACE IS FULLY WORN DUE TO REPETITIVE CONTACTS BETWEEN THE TWO BROKEN SECTIONS OF THE ROD. THE Ø3.2 PORTION PRESENTS SEVERAL MARKS COMING FROM THE SETSCREW AT ITS POSTERIOR SIDE AND SEVERAL CONTACTS WITH THE MAS CROWN AT ITS ANTERIOR SIDE (SOME MARKS AND CONTACTS APPEAR TWICE WHICH MAY CORRESPOND TO THE PRELIMINARY AND THE FINAL TIGHTENING). THE FRACTURE APPEARANCE IS TYPICAL OF FATIGUE DAMAGING OF THE MATERIAL WITH VISIBLE LINES OF REST ACROSS THE SECTION. THE FRACTURE APPEARANCE PRESENTS WORN SURFACE AT THE PERIPHERY OF THE SECTION DUE TO REPETITIVE CONTACTS BETWEEN THE TWO BROKEN SECTIONS OF THE ROD. NO PRE-EXISTING DEFECTS HAVE BEEN IDENTIFIED ON THE IMPLANTS THAT COULD BE RESPONSIBLE OF THE EVENT. THE RODS SHOWS TYPICAL METAL FATIGUE DAMAGING DUE TO OVERLOAD APPLIED ON THE SPINAL CONSTRUCT. THE FULL WORN SURFACE AT THE BROKEN SECTION OF THE LONGEST ROD DUE TO REPETITIVE CONTACTS BETWEEN THE TWO BROKEN SECTIONS OF THE ROD SUGGEST THAT THE INITIAL SURGERY HAS BEEN PERFORMED PRIOR THE SURGERY DATE REPORTED IN THE EVENT.
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN UNSPECIFIED SPINAL FUSION PROCEDURE WITH POSTERIOR FIXATION. IT WAS NOTED SOMETIME POST-OP THAT THE DEVICE HAD MIGRATED. A REVISION SURGERY REPORTEDLY TOOK PLACE 14 DAYS POST-OP. NO FURTHER INFORMATION IS KNOWN AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | VERTEX RECONSTRUCTION SYSTEM | APPLIANCE, FIXATION, SPINAL INTERLAMINAL | KWP | WARSAW ORTHOPEDICS, INC. | W05G3024 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 00066 YR |