ACCESS
Report
- Report Number
- 6000001-2011-08976
- Event Type
- Malfunction
- Date Received
- June 27, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 6, 2011
- Manufacturer
- BAXTER HEALTHCARE - MALTA
- Product Code
- FPA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- PHARMACIST
Narratives
(B)(4). AN ACTUAL SAMPLE WAS RECEIVED FOR EVALUATION. A VISUAL INSPECTION OF THE SAMPLE REVEALED THE Y-JUNCTION WAS DAMAGED. THE REPORTED CONDITION WAS CONFIRMED. ALTHOUGH THE CONDITION WAS CONFIRMED, THE ROOT CAUSE WAS NOT IDENTIFIED. A BATCH REVIEW WAS CONDUCTED AND NO ISSUES WERE FOUND RELATED TO THE REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH CAPA - (B)(4).
(B)(4). THE SAMPLE WAS RECEIVED FOR EVALUATION; HOWEVER, THE EVALUATION HAS NOT YET BEEN COMPLETED. A FOLLOW-UP REPORT WILL BE SUBMITTED UPON COMPLETION OF EVALUATION OR IF ADDITIONAL INFORMATION BECOMES AVAILABLE. THIS DEVICE IS MANUFACTURED FOR DISTRIBUTION OUTSIDE OF THE UNITED STATES (US); THEREFORE, IT DOES NOT HAVE A US 510K NUMBER. HOWEVER, THIS MDR IS BEING SUBMITTED BECAUSE IT IS THE SAME AS OR SIMILAR TO A PRODUCT DISTRIBUTED WITHIN THE US.
THE CUSTOMER REPORTED TO BAXTER (B)(4) A SOLUTION ADMINISTRATION SET IN WHICH LEAK OCCURRED. ACCORDING TO THE REPORT, AT THE BEGINNING OF ADMINISTRATION OF AN ANALGESIC SOLUTION, A BLOOD BACK FLOW OCCURRED RESULTING IN A LEAKAGE OF BLOOD MIXED WITH SOLUTION AT THE LEVEL OF THE Y-SITE. REPORTEDLY, IT IS DUE TO A DEFECTIVE Y-SITE WHICH APPEARS TO BE CRUSHED. THE CONDITION OCCURRED DURING INFUSION ON A PATIENT. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ACCESS | SET, ADMINISTRATION, INTRAVASCULAR | FPA | BAXTER HEALTHCARE - MALTA | 11B22V474 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNKNOWN ANALGESIC |