COLLEAGUE PRE P1.7
Report
- Report Number
- 6000001-2011-08971
- Event Type
- Malfunction
- Date Received
- June 27, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 16, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY WAS CONFIRMED AND DUPLICATED BY BAXTER SERVICE PERSONNEL. THE ROOT CAUSE WAS DETERMINED TO BE DEPLETED MAIN BATTERIES AS A RESULT OF USER ERROR. THE MAIN BATTERIES AND HARNESS WERE REPLACED TO CORRECT THIS CONDITION. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 5.09.90. ADDITIONAL INFORMATION: A DEVICE HISTORY REVIEW REVEALED NO ABNORMALITIES WERE FOUND DURING MANUFACTURING. A SERVICE HISTORY REVIEW REVEALED THAT THIS DEVICE WAS PREVIOUSLY SERVICED FOR THE REPORTED CONDITION. SHOULD ADDITIONAL INFORMATION BE RECEIVED, A FOLLOW-UP MEDWATCH WILL BE SUBMITTED.
(B)(4).THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
THE FACILITY REPRESENTATIVE REPORTED A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY. THIS CONDITION WAS FOUND UPON POWER-UP OF THE DEVICE, BUT IT IS UNKNOWN IN WHICH CARE AREA IT OCCURRED. THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. THE FACILITY REPRESENTATIVE STATED THAT THERE WAS NO PATIENT INVOLVEMENT, PATIENT INJURY OR MEDICAL INTERVENTION. NO ADDITIONAL INFORMATION IS AVAILABLE. THE USER INTERFACE MODULE SOFTWARE VERSION IS UNKNOWN AT THIS TIME.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE PRE P1.7 | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |