FDA Adverse Event Malfunction Summary report: N

RELIANCE 777 WASHER

MDR report key: 2141617 · Received June 27, 2011

Report

Report Number
9680353-2011-00017
Event Type
Malfunction
Date Received
June 27, 2011
Date of Event
June 1, 2011
Report Date
June 27, 2011
Manufacturer
STERIS CANADA CORPORATION
Product Code
LDS
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A STERIS TECHNICIAN INSPECTED THE WASHER AND FOUND THE INDEXING MECHANISM WAS NOT WORKING PROPERLY DUE TO A BROKEN SPRING AND WORN INDEXING GUIDES; THIS RESULTED IN THE BASKET SHIFTING ON THE CONVEYOR AND BECOMING STUCK. THE TECHNICIAN REPLACED THE SPRING AND GUIDES AND PLACED THE UNIT BACK INTO SERVICE. NO FURTHER ISSUES HAVE BEEN REPORTED WITH THE EQUIPMENT. THE UNIT IS UNDER STERIS SERVICE CONTRACT AND WAS LAST SERVICED ON (B)(4) 2011. THE STERIS TECHNICIAN CONFIRMED DURING THIS VISIT THE WASHER INDEXING MECHANISM WAS OPERATING PROPERLY; NO ISSUES WERE NOTED.

Description of Event or Problem · 1

THE USER FACILITY REPORTED THAT WHEN AN OPERATOR OPENED THE SERVICE ACCESS DOOR AND ATTEMPTED TO REMOVE A JAMMED BASKET AFTER THE WASHER ALARMED, THE BASKET POPPED UP AND HIT THE OPERATOR'S HAND. THE OPERATOR OBTAINED A BRUISE AND WENT TO EMPLOYEE HEALTH; THE USER FACILITY WOULD NOT PROVIDE ADDITIONAL INJURY OR TREATMENT INFORMATION. THE USER FACILITY REPORTED THE OPERATOR IS FINE AND HAS NO SUSTAINING INJURIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RELIANCE 777 WASHER WASHER LDS STERIS CANADA CORPORATION

Patients

Seq Age Sex Outcome Treatment
1