PACKAGE, VISUM LED2 SURGICAL LIGHT, CAMERA READY
Report
- Report Number
- 2031963-2011-00041
- Event Type
- Malfunction
- Date Received
- June 1, 2011
- Date of Event
- May 4, 2011
- Report Date
- May 4, 2011
- Manufacturer
- STRYKER COMMUNICATIONS
- Product Code
- FSY
- PMA / PMN Number
- K060802
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- EI
- Reporter Occupation
- NOT APPLICABLE
Narratives
THERE WAS NO PT INVOLVEMENT, THUS NO PT DATA EXISTS. THERE IS NO EXPIRATION DATE FOR THIS PRODUCT. THE INFO IN THIS MEDWATCH WAS SUPPLIED TO STRYKER (B)(4) BY STRYKER'S (B)(4) IN (B)(4) DURING INSTALLATION. THE DEVICE WAS INSPECTED ON SITE DURING INSTALLATION. EVAL SUMMARY: IT WAS REPORTED THAT DURING INSTALLATION OF A LIGHT HEAD, IT WAS DISCOVERED THE LIGHT WAS MISSING THE CIRCLIP. THE CIRCLIP IS USED TO HOLD THE LIGHT HEAD TO THE CARDANIC. IF THE CLIP IS MISSING, THE LIGHT HEAD COULD POTENTIALLY FALL. IF THIS WERE TO OCCUR DURING A CASE, THIS COULD HAVE RESULTED IN A SERIOUS OR SEVERE ADVERSE HEALTH EVENT. THIS TYPE OF EVENT DID NOT OCCUR FOR THIS INSTANCE BUT COULD HAVE OCCURRED. A MISSING CIRCLIP ON A LIGHT HEAD COULD RESULT IN THE LIGHT FALLING. IF THIS WERE TO OCCUR DURING CASE, THIS COULD HAVE RESULTED IN A SERIOUS OR SEVERE ADVERSE HEALTH EVENT. THIS EVENT OCCURRED DURING INSTALLATION, THUS, THERE WAS NO PT INVOLVEMENT OR INJURY. THIS INCIDENT WILL HAVE AN NCR OPENED AND WILL BE MONITORED THROUGH THE NCMB PROCESS. THIS IS NOT A SINGLE USE DEVICE.
(B)(4). (B)(6) REPORTED THAT IS A LOOSE SCREW INSIDE THE LIGHT HEAD AND ALSO A MISSING CIRCLIP THAT SHOULD RETAIN THE LIGHT HEAD TO THE SUSPENSION. (B)(6) FURTHER REPORTED THAT THE INSTALLATION AT THIS ACCOUNT IS NOT YET COMPLETE BUT HE EXPECTS THE ROOM TO OPEN WITHIN THE (B)(6). NON ADVERSE CONSEQUENCES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PACKAGE, VISUM LED2 SURGICAL LIGHT, CAMERA READY | FSY | STRYKER COMMUNICATIONS | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |