FDA Adverse Event Malfunction Summary report: N

AEM SUCTION IRRIGATION ELECTRODE, SPATULA

MDR report key: 2141284 · Received May 26, 2011

Report

Report Number
1722040-2011-00004
Event Type
Malfunction
Date Received
May 26, 2011
Date of Event
May 1, 2011
Report Date
May 27, 2011
Manufacturer
ENCISION
Product Code
GEI
PMA / PMN Number
K100711
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

HOSPITAL CONFIRMED THAT THE TIP CASE OFF THE SHAFT OF THE ELECTRODE DURING CLEANING IN SPD. NO PT WAS INVOLVED. THE HOSPITAL ESTIMATED THAT THEY HAD USED THE INSTRUMENT APPROXIMATELY 24 TIMES. THE MAXIMUM NUMBER OF USES, AS NOTED IN THE INSTRUCTIONS FOR USE, IS 25 TIMES. ACTUAL NUMBER OF USES CAN NOT BE DETERMINED WITH CERTAINTY. ENGINEERING ANALYSIS FOUND SMALL HOLES WHERE THE TIP BROKE AWAY AT THE WELD SITES. THE WELD ITSELF HELD AND THE TUBE TORE AWAY DUE TO FATIGUE OR EXCESSIVE FORCE.

Description of Event or Problem · 1

TIP BROKE OFF OF SUCTION IRRIGATION ELECTRODE WHILE BEING CLEANED IN STERILE PROCESSING DEPARTMENT. ITEMS ARE CLEANED BY HAND. NO PT INVOLVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AEM SUCTION IRRIGATION ELECTRODE, SPATULA ELECTROSURGICAL INSTRUMENT GEI ENCISION ES3882 QI

Patients

Seq Age Sex Outcome Treatment
1