FDA Adverse Event
Malfunction
Summary report: N
AEM SUCTION IRRIGATION ELECTRODE, SPATULA
MDR report key: 2141284
·
Received May 26, 2011
Report
- Report Number
- 1722040-2011-00004
- Event Type
- Malfunction
- Date Received
- May 26, 2011
- Date of Event
- May 1, 2011
- Report Date
- May 27, 2011
- Manufacturer
- ENCISION
- Product Code
- GEI
- PMA / PMN Number
- K100711
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
HOSPITAL CONFIRMED THAT THE TIP CASE OFF THE SHAFT OF THE ELECTRODE DURING CLEANING IN SPD. NO PT WAS INVOLVED. THE HOSPITAL ESTIMATED THAT THEY HAD USED THE INSTRUMENT APPROXIMATELY 24 TIMES. THE MAXIMUM NUMBER OF USES, AS NOTED IN THE INSTRUCTIONS FOR USE, IS 25 TIMES. ACTUAL NUMBER OF USES CAN NOT BE DETERMINED WITH CERTAINTY. ENGINEERING ANALYSIS FOUND SMALL HOLES WHERE THE TIP BROKE AWAY AT THE WELD SITES. THE WELD ITSELF HELD AND THE TUBE TORE AWAY DUE TO FATIGUE OR EXCESSIVE FORCE.
Description of Event or Problem · 1
TIP BROKE OFF OF SUCTION IRRIGATION ELECTRODE WHILE BEING CLEANED IN STERILE PROCESSING DEPARTMENT. ITEMS ARE CLEANED BY HAND. NO PT INVOLVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AEM SUCTION IRRIGATION ELECTRODE, SPATULA | ELECTROSURGICAL INSTRUMENT | GEI | ENCISION | ES3882 | QI |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |