OLYMPUS EVIS EXERA II BRONCHOVIDEOSCOPE
Report
- Report Number
- 8010047-2011-00125
- Event Type
- Other
- Date Received
- June 15, 2011
- Date of Event
- April 28, 2011
- Report Date
- May 17, 2011
- Manufacturer
- OLYMPUS MEDICAL SYSTEM CORPORATION
- Product Code
- EOQ
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MN, US
- Reporter Occupation
- OTHER
Narratives
THE DEVICE REFERENCED IN THIS REPORT WAS FORWARDED TO AN INDEPENDENT MICROBIOLOGY LABORATORY FOR MICROBIOLOGICAL TESTING. TEST RESULTS FROM THIS TESTING ARE PENDING. OLYMPUS WAS INFORMED THAT THREE OF THE FIVE REPORTED INFECTIONS TOOK PLACE IN (B)(4) 2011, ANOTHER ON (B)(4) 2011, AND THE FIFTH ON (B)(4) 2011. AS PART OF THE INVESTIGATION INTO THIS REPORT, AN OLYMPUS ENDOSCOPY SUPPORT SPECIALIST (ESS) VISITED THE USER FACILITY TO ASSESS THE REPROCESSING PRACTICES BEING EMPLOYED. THE USER FACILITY STAFF WERE NOT OBSERVED TO PERFORM PRE-CLEANING, AND A FLUSHING PUMP WAS REPORTEDLY NOT MAINTAINED IN ACCORDANCE WITH THE DIRECTIONS FOR USE. THE ESS HAS PROVIDED INFO REGARDING APPROPRIATE REPROCESSING AND MAINTENANCE OF THE BRONCHOSCOPES AND FLUSHING PUMP. A PHYSICAL EVALUATION OF THE BRONCHOSCOPE WILL FOLLOW ONCE THE DEVICE IS RECEIVED FROM THE MICROBIOLOGY LABORATORY, AND THIS REPORT WILL BE SUPPLEMENTED. THE CAUSE OF THE USER'S REPORT CANNOT BE CONCLUSIVELY DETERMINED. THIS REPORT IS BEING SUBMITTED AS AN MDR IN AN ABUNDANCE OF CAUTION. THIS IS REPORT TWO OF FIVE REPORTS: CROSS REFERENCE MFR. NUMBER 8010047-2011-00122, 8010047-2011-00123, 8010047-2011-00124, AND 8010047-2011-00126.
THE USER FACILITY REPORTED THAT FIVE PTS HAD ACQUIRED HEMOPHILUS HAEMOLYTICUS AND/OR MYCOBACTERIUM AVIUM COMPLEX INFECTIONS AFTER HAVING UNDERGONE A BRONCHOSCOPY WITH THE SUBJECT DEVICE. THE EXAMINATIONS WERE SAID TO HAVE OCCURRED BETWEEN (B)(6) 2011. ALL PTS WERE OUTPATIENTS. THE STAFF AT THE USER FACILITY FURTHER REPORTED THAT THEY DO NOT SUSPECT THE BRONCHOSCOPE TO BE THE CAUSE OF THE PTS' INFECTIONS. THERE WAS NO FURTHER INFO PROVIDED ON THE STATUS OF THE PT ASSOCIATED WITH THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | OLYMPUS EVIS EXERA II BRONCHOVIDEOSCOPE | BRONCHOSCOPE | EOQ | OLYMPUS MEDICAL SYSTEM CORPORATION | BF-Q180 | NA |