FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2140884 · Received June 27, 2011

Report

Report Number
2531779-2011-04485
Event Type
Malfunction
Date Received
June 27, 2011
Date of Event
May 27, 2011
Report Date
May 31, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VT, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/17/2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE WORN BUT INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT THE UP ARROW AND CONTRAST KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. IT WAS OBSERVED THAT THE UP ARROW BUTTON AND CONTRAST BUTTON DO NOT CLICK OR SPRING BACK NORMALLY WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT SHE NOTICED TWO DAYS AGO THAT THE UP ARROW KEYPAD BUTTON WAS FLAT AND INTERMITTENTLY UNRESPONSIVE. SHE DENIED EXPOSING THE PUMP TO CLEANING PRODUCTS, AND CONFIRMED THAT THERE WERE NO HOLES OR TEARS IN THE KEYPAD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 38 YR