FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2140884
·
Received June 27, 2011
Report
- Report Number
- 2531779-2011-04485
- Event Type
- Malfunction
- Date Received
- June 27, 2011
- Date of Event
- May 27, 2011
- Report Date
- May 31, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VT, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 06/17/2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE WORN BUT INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT THE UP ARROW AND CONTRAST KEYPAD BUTTONS ARE INTERMITTENTLY RESPONSIVE. IT WAS OBSERVED THAT THE UP ARROW BUTTON AND CONTRAST BUTTON DO NOT CLICK OR SPRING BACK NORMALLY WHEN PRESSED. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT SHE NOTICED TWO DAYS AGO THAT THE UP ARROW KEYPAD BUTTON WAS FLAT AND INTERMITTENTLY UNRESPONSIVE. SHE DENIED EXPOSING THE PUMP TO CLEANING PRODUCTS, AND CONFIRMED THAT THERE WERE NO HOLES OR TEARS IN THE KEYPAD.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 38 YR |