FDA Adverse Event Malfunction Summary report: N

CENTRA BED

MDR report key: 2140875 · Received May 31, 2011

Report

Report Number
1824206-2011-02976
Event Type
Malfunction
Date Received
May 31, 2011
Date of Event
May 10, 2011
Report Date
May 10, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SC, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

REPAIRS HAVE NOT BEEN COMPLETED.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE KNEE SECTION MOVED UP UNINTENTIONALLY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTRA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 850

Patients

Seq Age Sex Outcome Treatment
1