OMNIPOD INSULIN PUMP

MDR report key: 2140857 · Received May 31, 2011

Report

Report Number
3004464228-2011-00239
Event Type
Malfunction
Date Received
May 31, 2011
Date of Event
April 28, 2011
Report Date
April 28, 2011
Manufacturer
INSULET CORPORATION
Product Code
LZG
PMA / PMN Number
K042792
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE INVESTIGATION OF THE RETURNED POD FOUND DISCOLORATION INSIDE THE DEVICE, WHICH IS EVIDENCE OF AN INTERNAL FLUID LEAK. A LEAK TEST WAS PERFORMED, WHICH CONFIRMED THE PRESENCE OF A LEAK IN THE FLUID PATH CAUSED BY A TEAR IN THE CANNULA TUBING. THE LEAK WOULD HAVE RESULTED IN INSULIN COMING INTO CONTACT WITH INTERNAL COMPONENTS AND ASSEMBLIES, ULTIMATELY CAUSING THE DEVICE TO FAIL. A POD MALFUNCTION, THEREFORE, IS CONFIRMED TO HAVE BEEN A CONTRIBUTING FACTOR TO THE CUSTOMER'S HIGH BG LEVELS. A REVIEW OF LOT QUALIFICATION RECORDS REVEALED NO EVIDENCE OF THIS FAILURE MODE. LOT QUALIFICATION RESULTS WERE DEEMED ACCEPTABLE AND THE LOT PASSED THE ACCEPTANCE CRITERIA. INSULET HAS INITIATED AN INTERNAL INVESTIGATION INTO THIS PARTICULAR FAILURE MODE AND HAS TAKEN MULTIPLE ACTIONS TO MINIMIZE AND PREVENT ITS RECURRENCE. A RISK ASSESSMENT HAS ALSO BEEN CONDUCTED AS WELL AS ONGOING MONITORING TO ENSURE THAT THIS LEVEL OF FAILURE DOES NOT EXCEED ACTION LIMITS (AS DEFINED BY INSULET'S INTERNAL PROCEDURES). LABELING, TRAINING AND CLINICAL ADVICE CONTINUES TO INDICATE THAT ONGOING BG MONITORING IS NECESSARY. ALTHOUGH, WE DO NOT RELY ON LABELING, THIS IS ADDITIONAL MITIGATION AS PART OF THE NORMAL ACTIVITIES OF A DIABETIC.

Description of Event or Problem · 1

THE CUSTOMER CALLED TO REPORT THAT SHE FELT THE POD "HAD FAILED ON HER." THE DEVICE WAS APPLIED IN THE EVENING AND, BY EARLY MORNING, HER BG LEVELS HAD SIGNIFICANTLY INCREASED (FROM 104 TO 295MG/DL). A CORRECTION BOLUS WAS ADMINISTERED, BUT AN HOUR LATER, HER LEVELS CONTINUED TO RISE (UP TO 348MG/DL). THE POD WAS IMMEDIATELY DEACTIVATED AND WILL BE RETURNED FOR EVALUATION. NO SPECIFIC DEVICE FAILURE MODE WAS NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OMNIPOD INSULIN PUMP PUMP, INFUSION, INSULIN LZG INSULET CORPORATION 11200 L30467

Patients

Seq Age Sex Outcome Treatment
1 34 YR Other