FDA Adverse Event Malfunction Summary report: N

TEMP PUMP TEMPERATURE

MDR report key: 2140753 · Received June 15, 2011

Report

Report Number
1313850-2011-00144
Event Type
Malfunction
Date Received
June 15, 2011
Date of Event
May 20, 2011
Report Date
May 20, 2011
Manufacturer
STRYKER CORP DBA GAYMAR
Product Code
ILO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

RESULT: COLD SOLDER JOINT CAUSED TRACE ON PC BOARD TO FAIL. IT SHOULD BE NOTED THAT THE UNIT WAS RETURNED TO MFR IN NON-FUNCTIONING CONDITION.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THE BOARD WAS ARCING AND BURNING. NO PT INVOLVEMENT OR ADVERSE CONSEQUENCES WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 TEMP PUMP TEMPERATURE PUMP ILO STRYKER CORP DBA GAYMAR TP500 NA

Patients

Seq Age Sex Outcome Treatment
1