POSEY STAYSAFE BED
Report
- Report Number
- 2020362-2011-00217
- Event Type
- Malfunction
- Date Received
- June 15, 2011
- Report Date
- May 16, 2011
- Manufacturer
- J. T. POSEY CO.
- Product Code
- FNL
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS, US
- Reporter Occupation
- OTHER
Narratives
(B)(4) - ZIPPER TEETH WON'T CONNECT. EVALUATION RESULTS: EVALUATION OF THE PRODUCT FOUND THAT ON THE ACCESS WINDOW OF PANEL SIDE B, BOTH ZIPPER SLIDER BODIES ARE OPEN. WINDOW SIDE A HAS 1 INCH BROKEN STITCHES ON THE ZIPPER TAPE. WINDOW B HAS TWO HOLES IN THE NETTING. PANEL SIDE B INSERT PIN IS RIPPED FROM THE TOP RIDGE ZIPPER. NOTE: POSEY INSTRUCTIONS FOR USE HAS A WARNING STATEMENT: NEVER TRY TO RIP A PANEL OPEN AS THIS MAY DAMAGE THE ACCESS PANEL OR THE ZIPPER SLIDER BY BENDING IT OPEN. NEVER USE THE BED IF A ZIPPER SLIDER IS BENT OPEN OR DAMAGED, AS THIS MAY PREVENT THE ZIPPER FROM CLOSING SECURELY. NEVER LEAVE THE PATIENT'S BEDSIDE UNTIL ALL ACCESS PANEL ZIPPERS ARE SECURELY CLOSED. TEST THE ENTIRE LENGTH OF EACH ZIPPER BY PRESSING AGAINST THE PANEL NEAR THE ZIPPER TO MAKE SURE IT IS SECURELY CLOSED AND THE PANEL WILL NOT OPEN WHEN PRESSURE IS APPLIED. (B)(4).
THE CUSTOMER REPORTED THAT THE CANOPY HAS A ZIPPER SLIDER THAT IS BROKEN AND DOES NOT CONNECT THE TEETH ON PANEL SIDE A. THERE WAS NO PATIENT INCIDENT OR INJURY REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | POSEY STAYSAFE BED | FNL | J. T. POSEY CO. | 8060 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NI |