FDA Adverse Event Malfunction Summary report: N

POSEY STAYSAFE BED

MDR report key: 2140705 · Received June 15, 2011

Report

Report Number
2020362-2011-00217
Event Type
Malfunction
Date Received
June 15, 2011
Report Date
May 16, 2011
Manufacturer
J. T. POSEY CO.
Product Code
FNL
PMA / PMN Number
EXEMPT
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - ZIPPER TEETH WON'T CONNECT. EVALUATION RESULTS: EVALUATION OF THE PRODUCT FOUND THAT ON THE ACCESS WINDOW OF PANEL SIDE B, BOTH ZIPPER SLIDER BODIES ARE OPEN. WINDOW SIDE A HAS 1 INCH BROKEN STITCHES ON THE ZIPPER TAPE. WINDOW B HAS TWO HOLES IN THE NETTING. PANEL SIDE B INSERT PIN IS RIPPED FROM THE TOP RIDGE ZIPPER. NOTE: POSEY INSTRUCTIONS FOR USE HAS A WARNING STATEMENT: NEVER TRY TO RIP A PANEL OPEN AS THIS MAY DAMAGE THE ACCESS PANEL OR THE ZIPPER SLIDER BY BENDING IT OPEN. NEVER USE THE BED IF A ZIPPER SLIDER IS BENT OPEN OR DAMAGED, AS THIS MAY PREVENT THE ZIPPER FROM CLOSING SECURELY. NEVER LEAVE THE PATIENT'S BEDSIDE UNTIL ALL ACCESS PANEL ZIPPERS ARE SECURELY CLOSED. TEST THE ENTIRE LENGTH OF EACH ZIPPER BY PRESSING AGAINST THE PANEL NEAR THE ZIPPER TO MAKE SURE IT IS SECURELY CLOSED AND THE PANEL WILL NOT OPEN WHEN PRESSURE IS APPLIED. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE CANOPY HAS A ZIPPER SLIDER THAT IS BROKEN AND DOES NOT CONNECT THE TEETH ON PANEL SIDE A. THERE WAS NO PATIENT INCIDENT OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 POSEY STAYSAFE BED FNL J. T. POSEY CO. 8060 NA

Patients

Seq Age Sex Outcome Treatment
1 NI