FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2140682 · Received June 27, 2011

Report

Report Number
2531779-2011-04483
Event Type
Malfunction
Date Received
June 27, 2011
Date of Event
April 29, 2011
Report Date
May 30, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE WORN BUT INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT THE CONTRAST KEYPAD BUTTON IS INTERMITTENTLY RESPONSIVE; ALL OTHER KEYPAD BUTTONS WERE FOUND TO RESPOND APPROPRIATELY TO PRESSES. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE DOWN ARROW KEYPAD BUTTON BECAME INTERMITTENTLY UNRESPONSIVE A MONTH AGO. SHE STATED THAT THERE ARE NO HOLES OR TEARS IN THE KEYPAD; STATED THAT SHE WEARS THE PUMP IN HER POCKET; AND DENIED CLEANING THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 30 YR