FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2140682
·
Received June 27, 2011
Report
- Report Number
- 2531779-2011-04483
- Event Type
- Malfunction
- Date Received
- June 27, 2011
- Date of Event
- April 29, 2011
- Report Date
- May 30, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE WORN BUT INTACT; NO PEELING OR LIFTING WAS OBSERVED. EVALUATION REVEALED THAT THE CONTRAST KEYPAD BUTTON IS INTERMITTENTLY RESPONSIVE; ALL OTHER KEYPAD BUTTONS WERE FOUND TO RESPOND APPROPRIATELY TO PRESSES. THERE WAS EVIDENCE OF KEYPAD ADHESIVE FOUND UNDER ALL KEY CONTACTS.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT THE DOWN ARROW KEYPAD BUTTON BECAME INTERMITTENTLY UNRESPONSIVE A MONTH AGO. SHE STATED THAT THERE ARE NO HOLES OR TEARS IN THE KEYPAD; STATED THAT SHE WEARS THE PUMP IN HER POCKET; AND DENIED CLEANING THE PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 30 YR |