FDA Adverse Event
Malfunction
Summary report: N
ONETOUCHPING GLUCOSEMGMTSYSTEM
MDR report key: 2140662
·
Received June 27, 2011
Report
- Report Number
- 2531779-2011-04482
- Event Type
- Malfunction
- Date Received
- June 27, 2011
- Report Date
- May 31, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE DOWN AND OK BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE PUMP WAS OPERATING WITHIN SPECIFICATION AT THE TIME OF RELEASE.
Description of Event or Problem · 1
THE PATIENT REPORTED THAT THE DOWN AND OK BUTTONS ARE NOT RESPONDING CORRECTLY TO BUTTON PRESSES. THE PATIENT REPORTED THAT IT FEELS LIKE THE BUTTONS ARE STICKING AT TIMES. THE PATIENT STATED THAT THE KEYPAD IS INTACT BUT FEELS LOOSE. THE PATIENT REPORTEDLY WORE THE PUMP IN A CASE ON HER BELT AND DID NOT CLEAN THE PUMP.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |