FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 2140662 · Received June 27, 2011

Report

Report Number
2531779-2011-04482
Event Type
Malfunction
Date Received
June 27, 2011
Report Date
May 31, 2011
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2011 WITH THE FOLLOWING FINDINGS: THE DOWN AND OK BUTTONS WERE FOUND TO BE INTERMITTENTLY RESPONDING TO PRESSES. THE KEYPAD WAS REMOVED AND ADHESIVE WAS OBSERVED UNDER THE BUTTON CONTACTS. A REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THAT THE PUMP WAS OPERATING WITHIN SPECIFICATION AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PATIENT REPORTED THAT THE DOWN AND OK BUTTONS ARE NOT RESPONDING CORRECTLY TO BUTTON PRESSES. THE PATIENT REPORTED THAT IT FEELS LIKE THE BUTTONS ARE STICKING AT TIMES. THE PATIENT STATED THAT THE KEYPAD IS INTACT BUT FEELS LOOSE. THE PATIENT REPORTEDLY WORE THE PUMP IN A CASE ON HER BELT AND DID NOT CLEAN THE PUMP.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION ONETOUCH PING INSULIN PUMP

Patients

Seq Age Sex Outcome Treatment
1 29 YR