ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2011-04478
- Event Type
- Malfunction
- Date Received
- June 27, 2011
- Report Date
- May 31, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 (B)(6) 2011 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(6) 2011 WITH THE FOLLOWING FINDINGS: NO PHYSICAL DAMAGE WAS FOUND TO THE BATTERY COMPARTMENT OR BATTERY CAP. THE PUMP BOOTED TO THE VERIFY SCREEN WITH APPROPRIATE ALARMS. A TEST BOLUS WAS PERFORMED WITH NO POWER LOSS OCCURING. A REVIEW OF THE BLACK BOX SHOWED A REBOOT JUST AFTER AN OCCLUSION ALARM ON THE REPORTED EVENT DATE; THE PUMP WAS POWERED ON BACK ON FOR EIGHT MINUTES AND EMITTED 3 NO PRIME WARNINGS. THE PUMP WAS EXERCISED FOR 24 HOURS WITHOUT REBOOTING OR LOSING POWER. THE PUMP WAS OPENED AND THERE WERE NO DEFECTS TO THE POWER CIRCUIT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
THE PATIENT REPORTED THAT THE PUMP WILL NOT TURN ON. THE PATIENT RECEIVED A CS ALARM A FEW DAYS AGO, SHE REBOOTED THE PUMP AND IT WORKED FINE. WHEN SHE WAS GIVING A BOLUS SHE RECEIVED AN OCCLUSION ALARM AND THEN THE POWER WAS OFF TO THE PUMP. SHE CHANGED THE BATTERY THREE TIMES AND THE PUMP WILL NOT TURN ON.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ONETOUCH PING INSULIN PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR |