PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM
Report
- Report Number
- 2134265-2011-02425
- Event Type
- Malfunction
- Date Received
- June 27, 2011
- Date of Event
- May 30, 2011
- Report Date
- May 31, 2011
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE IS COMBINATION PRODUCT. (B)(4).
DEVICE EVALUATED BY MFR.: A VISUAL AND MICROSCOPIC EXAMINATION OF THE DEVICE FOUND THE DEVICE WAS RETURNED WITH THE STENT MOVED 3 MM DISTAL TO THE DISTAL MARKERBAND. STRUT ROWS TOWARDS THE MIDDLE END OF THE STENT WERE DAMAGED AND COMPRESSED LONGITUDINALLY. THE OUTER DIAMETER (OD) OF THE STENT DAMAGE WAS APPROXIMATELY 1.92 MM. THE OD OF THE STENT AREA WHERE NO DAMAGE WAS PRESENT WAS MEASURED AT APPROXIMATELY 1.23 MM. THE STENT LENGTH WAS MEASURED AT APPROXIMATELY 20 MM. THERE WERE ALSO KINKS IDENTIFIED ALONG THE ENTIRE LENGTH OF THE HYPOTUBE SHAFT. THE TIP AND BALLOON SECTIONS OF THE DEVICE WERE VISUALLY AND MICROSCOPICALLY EXAMINED AND NO ISSUES WERE NOTED WITH THEIR PROFILES THAT COULD HAVE POTENTIALLY CONTRIBUTED TO THE COMPLAINT INCIDENT. THE BALLOON WAS TIGHTLY WRAPPED AND WAS NOT SUBJECTED TO POSITIVE PRESSURE. A 0.015 INCH PRODUCT MANDREL WAS INSERTED THROUGH THE LUMEN WITH NO RESTRICTIONS NOTED. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS CONSIDERED HANDLING DAMAGE AS THE EVENT OCCURRED PRIOR TO PATIENT CONTACT. (B)(4).
IT WAS REPORTED THAT DURING PREPARATION FOR A STENTING TREATMENT PROCEDURE STENT DAMAGE OCCURRED. WHILE REMOVING THE 3.50MM X 24MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS) FROM IT'S PACKAGING, IT WAS NOTED THAT THE STENT STRUTS WERE DAMAGED. THE DEVICE DID NOT ENTER THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CURRENT CONDITION IS STABLE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE.
IT WAS REPORTED THAT DURING PREPARATION FOR A STENTING TREATMENT PROCEDURE STENT DAMAGE OCCURRED. WHILE REMOVING THE 3.50MM X 24MM PROMUS ELEMENT STENT DELIVERY SYSTEM (SDS) FROM IT'S PACKAGING, IT WAS NOTED THAT THE STENT STRUTS WERE DAMAGED. THE DEVICE DID NOT ENTER THE PATIENT AND THE PROCEDURE WAS COMPLETED WITH ANOTHER OF THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S CURRENT CONDITION IS STABLE. THIS PRODUCT IS ONLY OUS APPROVED BUT IT IS SIMILAR TO AN APPROVED US DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PROMUS ELEMENT EVEROLIMUS-ELUTING CORONARY STENT SYSTEM | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911324350 | 0014044581 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 80 YR |