FDA Adverse Event Malfunction Summary report: N

ETS-FLEX - ENDOSCOPIC

MDR report key: 2140506 · Received June 27, 2011

Report

Report Number
3005075853-2011-02594
Event Type
Malfunction
Date Received
June 27, 2011
Date of Event
May 31, 2011
Report Date
June 2, 2011
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K020779
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ANALYSIS RESULTS FOUND THAT THE ATW35 DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH NO RELOAD PRESENT ON THE DEVICE. THE DEVICE WAS TESTED FOR FUNCTIONALITY WITH A TEST RELOAD AND IT FIRED, CUT AND FORMED THE STAPLES AS INTENDED. THE DEVICE FIRED WITHOUT ANY DIFFICULTIES, THE STAPLE LINE WAS COMPLETE, THE CUT LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. EVENT COULD NOT BE CONFIRMED AS THE DEVICE OPENED AND CLOSED WITHOUT ANY DIFFICULTIES NOTED.

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A ZANKERS DIVERTICULITIS PROCEDURE, THAT THE FIRING TRIGGER WAS STUCK IN THE "FIRED" POSITION AND WOULD NOT REOPEN. THERE WERE NO PATIENT CONSEQUENCES. UNKNOWN HOW CASE WAS COMPLETED.