PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2011-81815
- Event Type
- Injury
- Date Received
- June 16, 2011
- Date of Event
- June 7, 2011
- Report Date
- June 7, 2011
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS MED-REL
- Product Code
- MDS
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
INSULIN PUMP RECEIVED IN WITH NO BUTTON RESPONSE DUE TO CORRODED KEYPAD TRACES. UNABLE TO PERFORM THE PRIME, DISPLACEMENT, BASIC OCCLUSION, OCCLUSION AND EXCESSIVE NO DELIVERY TESTS DUE TO THE KEYPAD ANOMALY. A CRACKED CASE, CRACKED RESERVOIR TUBE LIP, PEELING KEYPAD OVERLAY AND A SCRATCHED DISPLAY WINDOW NOTED DURING VISUAL INSPECTION.
CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. WE THEREFORE CONSIDER THIS REPORT COMPLETE TO THE BEST OF OUR KNOWLEDGE.
IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED FOR HIGH BLOOD GLUCOSE. IT WAS STATED THAT THE INSULIN PUMP HAD A BUTTON ERROR ALARM. TROUBLESHOOTING WAS PERFORMED. THE BATTERY WAS CHANGED AND THE INSULIN PUMP WAS NOT RESPONDING. NO FURTHER INFO WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | MDS | MEDTRONIC PUERTO RICO OPERATIONS MED-REL | MMT-722WWP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization |