FDA Adverse Event Death Summary report: N

VOLUSON E8

MDR report key: 2140423 · Received June 21, 2011

Report

Report Number
8020021-2011-00003
Event Type
Death
Date Received
June 21, 2011
Date of Event
May 12, 2011
Report Date
June 21, 2011
Manufacturer
GE HEALTHCARE
Product Code
IYO
PMA / PMN Number
K061682
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DATE OF INTRAUTERINE DEATH IS UNDETERMINED, BUT WAS DETECTED (B)(4) 2011. THE DEVICE WAS EVALUATED BY GE HEALTHCARE ENGINEERING AND THE SYS WAS DETERMINED TO BE OPERATING PER SPECIFICATIONS. THE PT'S DATA FILE FROM (B)(4) 2011 WAS AVAILABLE AND ANALYZED. THE MEASUREMENT FUNCTIONALITY OF THE ULTRASOUND SYS WAS ANALYZED AND THE RESULTS WERE VERIFIED AGAINST THE SCALE ON THE SIDE OF THE SCREEN. NO ABNORMALITY WAS DETECTED. THE DEVICE'S CALCULATIONS AND PERCENTILES WERE VERIFIED WITH PUBLISHED GROWTH CHARTS. NO ABNORMALITY WAS DETECTED. THE ESTIMATED FETAL WEIGHT CALCULATION WAS VERIFIED WITH THE PUBLISHED CHART AND AGAIN NO ABNORMALITY WAS DETECTED. BASED ON THE EVAL OF THE SYS AND THE RE-ANALYSIS OF THE (B)(4) 2011 DATA, THE SYS PERFORMED PER SPECIFICATIONS. THE CAUSE OF THE INTRAUTERINE DEATH IS UNDETERMINED OR HAS NOT BEEN MADE AVAILABLE TO GE HEALTHCARE. NO FURTHER ACTIONS ARE PLANNED AT THIS TIME.

Description of Event or Problem · 1

AN OBSTETRIC PT PRESENTED ON (B)(6) 2011, AT A GESTATIONAL AGE OF 38+6 WEEKS. DURING ROUTINE MEASUREMENTS, THE SONOGRAPHER CALCULATED AN ESTIMATED FETAL WEIGHT (EFW) OF 7 LB 13 OZ, THE BIPARIETAL DIAMETER (BPD) SHOWING UP AT THE 18%, THE HEAD CIRCUMFERENCE (HC) BELOW 2.3%, ABDOMINAL CIRCUMFERENCE (AC) AT 95%, AND FEMUR LENGTH (FL) AT 12%. THE PT RETURNED ON (B)(6) 2011 AND AN INTRAUTERINE DEATH WAS DETECTED. THE FETUS WAS DELIVERED AND THE WEIGHT WAS APPROX 10 LB. THE HEAD CIRCUMFERENCE ALSO APPEARED LARGE AFTER DELIVERY ALTHOUGH AN ACTUAL MEASUREMENT WAS NOT AVAILABLE. THE DOCTOR INDICATED THAT IF HE HAD KNOWN ABOUT THE TRUE WEIGHT OF THE FETUS, HE WOULD HAVE TAKEN THE MOTHER'S BLOOD GLUCOSE TO MONITOR HER TREATMENT FOR DIABETES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VOLUSON E8 ULTRASONIC PULSED ECHO IMAGING SYSTEM IYO GE HEALTHCARE

Patients

Seq Age Sex Outcome Treatment
1 29 YR Death