PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 2140416 · Received June 16, 2011

Report

Report Number
3004209178-2011-81836
Event Type
Injury
Date Received
June 16, 2011
Date of Event
June 5, 2011
Report Date
June 5, 2011
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS MED-REL
Product Code
MDS
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNK WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. FURTHER INFO WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED FOR DIABETES KETOACIDOSIS AND HEADACHES. THE BLOOD GLUCOSE READING WAS 500 MG/DL. TROUBLESHOOTING WAS PERFORMED. THE PROGRAMMING, TIME, AND DATE ON THE INSULIN PUMP WERE CORRECT. THE ALARM HISTORY REVEALED SEVERAL LOW BATTERY ALARMS. THE MOTHER STATED THAT THE CUSTOMER JUMPED INTO THE HOT TUB WHILE WEARING THE INSULIN PUMP. RAN A FIXED PRIME AND HIGH PRESSURE TEST AND THE TESTS PASSED. NO FURTHER INFO WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP MDS MEDTRONIC PUERTO RICO OPERATIONS MED-REL MMT-723NAB

Patients

Seq Age Sex Outcome Treatment
1 18 YR Hospitalization