FDA Adverse Event Malfunction Summary report: N

MICRO DRILL MEDIUM STRAIGHT ATTACHMENT

MDR report key: 2140370 · Received June 17, 2011

Report

Report Number
1811755-2011-02236
Event Type
Malfunction
Date Received
June 17, 2011
Date of Event
May 24, 2011
Report Date
May 28, 2011
Manufacturer
STRYKER INSTRUMENTS KALAMAZOO
Product Code
ERL
PMA / PMN Number
K040300
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE PROBABLE CAUSE OF THE BUR BREAKING WAS ATTRIBUTED TO DEBRIS IN THE BUR PILOT WHICH CAUSED THE WEDGE COLLET TO FAIL TO OPEN AND RELEASE THE BUR. THE MICRO DRILL MEDIUM STRAIGHT ATTACHMENT IS NOT A REPAIRABLE DEVICE; IT WAS DISCARDED AFTER THE QUALITY INVESTIGATION WAS COMPLETE.

Description of Event or Problem · 1

THE MICRO DRILL MEDIUM STRAIGHT ATTACHMENT WAS SENT IN FOR EVAL BECAUSE IT HAS A BROKEN BUR IN IT. THE EVENT WAS NOTICED PRIOR TO A PROCEDURE; NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MICRO DRILL MEDIUM STRAIGHT ATTACHMENT ERL STRYKER INSTRUMENTS KALAMAZOO 10130

Patients

Seq Age Sex Outcome Treatment
1 UNK