FDA Adverse Event
Malfunction
Summary report: N
MICRO DRILL MEDIUM STRAIGHT ATTACHMENT
MDR report key: 2140370
·
Received June 17, 2011
Report
- Report Number
- 1811755-2011-02236
- Event Type
- Malfunction
- Date Received
- June 17, 2011
- Date of Event
- May 24, 2011
- Report Date
- May 28, 2011
- Manufacturer
- STRYKER INSTRUMENTS KALAMAZOO
- Product Code
- ERL
- PMA / PMN Number
- K040300
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
THE PROBABLE CAUSE OF THE BUR BREAKING WAS ATTRIBUTED TO DEBRIS IN THE BUR PILOT WHICH CAUSED THE WEDGE COLLET TO FAIL TO OPEN AND RELEASE THE BUR. THE MICRO DRILL MEDIUM STRAIGHT ATTACHMENT IS NOT A REPAIRABLE DEVICE; IT WAS DISCARDED AFTER THE QUALITY INVESTIGATION WAS COMPLETE.
Description of Event or Problem · 1
THE MICRO DRILL MEDIUM STRAIGHT ATTACHMENT WAS SENT IN FOR EVAL BECAUSE IT HAS A BROKEN BUR IN IT. THE EVENT WAS NOTICED PRIOR TO A PROCEDURE; NO ADVERSE EVENT WAS ASSOCIATED WITH THE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MICRO DRILL MEDIUM STRAIGHT ATTACHMENT | ERL | STRYKER INSTRUMENTS KALAMAZOO | 10130 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK |