ZENITH AAA ENDOVASCULAR GRAFT ILIAC LEG
Report
- Report Number
- 1820334-2011-00357
- Event Type
- Injury
- Date Received
- June 16, 2011
- Report Date
- June 9, 2011
- Manufacturer
- COOK, INC.
- Product Code
- MIH
- PMA / PMN Number
- P020018
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PHYSICIAN
Narratives
(B)(4) - CLAUDICATION IS LABELED IN THE IFU. STENT DISCONNECTING FROM GRAFT IS NOT LABELED IN THE IFU. EVENT EVALUATION: STILL UNDER INVESTIGATION.
A (B)(6) YEAR OLD MALE PATIENT UNDERWENT INITIAL ENDOVASCULAR AORTIC REPAIR ON (B)(6), 2004. DURING FOLLOW-UP, A COMPLETE DISCONNECT OF GRAFT FROM SUPRARENAL STENT WAS OBSERVED AND THE GRAFT HAD FALLEN DOWN INTO THE SAC. INFORMATION WAS PROVIDED TO THE SALES REP THAT THE PATIENT HAD COME ANNUALLY FOR FOLLOW-UP. A REVIEW OF THE 2010 IMAGING INDICATES IT CLEARLY HAD SOME BREAKS AT THAT POINT, BUT HAD GONE UNREPORTED BY THE RADIOLOGIST READING THE SCAN. THE SUPRARENAL STENT APPEARED TO HAVE SUTURE BREAKS AND THE GRAFT WAS DISCONNECTING FROM THE STENT ON THE LEFT SIDE OF THE AORTIC NECK, STILL APPOSED AND HAD NOT DROPPED AT THIS TIME. THE PATIENT PRESENTED WITH BUTTOCK/LEG CLAUDICATION AND HIS FOLLOW UP 2011 SCAN REVEALED A COMPLETE DISCONNECT. THE PATIENT WILL BE BROUGHT BACK TO THE OPERATING ROOM FOR POSSIBLE SECONDARY INTERVENTION. THE PHYSICIAN WILL SEE IF THEY CAN GET STIFF WIRE ACCESS AND STRAIGHTEN THE GRAFT IN AN EFFORT TO REALIGN AND COVER THE EXPOSED SEGMENT. MIGRATED GRAFT, BUTTOCK/LEG CLAUDICATION, MINIMAL PEDAL PULSES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ZENITH AAA ENDOVASCULAR GRAFT ILIAC LEG | MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | COOK, INC. | NA | UNKNOWN |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 59 YR | Other |