FDA Adverse Event Malfunction Summary report: N

CENTRA BED

MDR report key: 2140350 · Received June 16, 2011

Report

Report Number
1824206-2011-03294
Event Type
Malfunction
Date Received
June 16, 2011
Date of Event
May 26, 2011
Report Date
May 26, 2011
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECH FOUND THAT TOP BLOCK WAS CRACKED AND THE CASTERS WERE WORN. HE REPLACED TOP BLOCK, INSTALLED NEW BEARING AND REPLACED THE CASTERS TO REPAIR THE BED.

Description of Event or Problem · 1

INFO RECEIVED INDICATES THE BRAKES APPEAR TO BE SET BUT THE CASTER WHEELS WILL STILL ROLL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTRA BED AC-POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 852

Patients

Seq Age Sex Outcome Treatment
1