FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3100 SYSTEM

MDR report key: 2140339 · Received June 15, 2011

Report

Report Number
3008642652-2011-00056
Event Type
Malfunction
Date Received
June 15, 2011
Date of Event
May 12, 2011
Report Date
June 14, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DEVICE IS ALARMING) HAS BEEN CONFIRMED. UPON INVESTIGATION, THE CABLE RUNNING FROM THE DISTRIBUTION NODE TO ECG B HAS AN OPEN INTERNAL WIRE. THE ROOT CAUSE OF THE OPEN INTERNAL WIRE CANNOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE OPEN INTERNAL WIRE. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) FEMALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE DEVICE IS ALARMING WITH BOTH GARMENTS ON. THE PATIENT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 3100 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3000 NA

Patients

Seq Age Sex Outcome Treatment
1 58 YR