FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 2140311 · Received June 15, 2011

Report

Report Number
3008642652-2011-00051
Event Type
Malfunction
Date Received
June 15, 2011
Date of Event
May 2, 2011
Report Date
June 14, 2011
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CONNECTOR WILL NOT STAY LATCHED) HAS BEEN CONFIRMED. UPON EVALUATION, THE TRUNK CABLE CONNECTOR WAS DAMAGED. THE CAUSE FOR THE DAMAGED CONNECTOR CANNOT BE POSITIVELY DETERMINED BUT WAS LIKELY THE RESULT OF EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE ELECTRODE BELT CONNECTOR. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) FEMALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT CONSTANT CHECK BELT MESSAGES AND THAT THE CONNECTION BETWEEN HER MONITOR AND ELECTRODE BELT WAS LOOSE. THE PATIENT WAS PROVIDED WITH A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 57 YR