FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 2140302
·
Received June 15, 2011
Report
- Report Number
- 3008642652-2011-00049
- Event Type
- Malfunction
- Date Received
- June 15, 2011
- Date of Event
- April 4, 2011
- Report Date
- June 14, 2011
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF MONITOR (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WON'T POWER UP) HAS BEEN CONFIRMED. UPON EVALUATION, THERE WAS EVIDENCE THAT THE MONITOR EXPERIENCED A FAILURE WHICH DAMAGED MULTIPLE COMPONENTS ON THE COMPUTER/ANALOG BOARD AND THE DEFIBRILLATOR BOARD. THE CAUSE OF THE FAILURE CANNOT BE POSITIVELY DETERMINED. NO ADVERSE EVENT RESULTED FROM THE DAMAGED COMPONENTS. THE PATIENT RECEIVED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
A (B)(6) FEMALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT NEITHER OF HER BATTERIES WOULD POWER UP HER MONITOR. THE PATIENT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 25 YR |