FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600 PRO SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 2140276
·
Received June 27, 2011
Report
- Report Number
- 2050012-2011-02648
- Event Type
- Malfunction
- Date Received
- June 27, 2011
- Date of Event
- May 27, 2011
- Report Date
- May 27, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CTS ASSISTED THE CUSTOMER IN TROUBLESHOOTING AND FOUND THAT THE REAGENT SYRINGE WAS LOOSE AND FLUID HAD PUDDLED ON THE REAGENT TRAY. CTS DIRECTED THE CUSTOMER TO RESEAT THE SYRINGE BARREL AND CLEAN LIQUID. CTS DIRECTED CUSTOMER TO CHECK THE MIXER PADDLES AND THE CUSTOMER DISCOVERED THE REAGENT PADDLE WAS MISSING; CUSTOMER DID NOT HAVE A SPARE. A BCI FIELD SERVICE ENGINEER (FSE) FOUND REAGENT PROBE LEAKING AND CUVETTE NOT DRY BEFORE REAGENT INJECT. FSE UPDATED CHEMISTRY SOFTWARE FOR CUSTOMER AND RAN CONTROLS. REPAIRS VERIFIED PER ESTABLISHED PROCEDURES.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) CUSTOMER TECHNICAL SUPPORT (CTS) TO REPORT FLUID LEAK FROM THE REAGENT PROBE. NO INJURY WAS REPORTED OR OCCURRED DUE TO THIS EVENT.