FDA Adverse Event Summary report: N

UNICEL® DXC 800 SYNCHRON ® CHEMISTRY ANALYZER

MDR report key: 2140237 · Received June 26, 2011

Report

Report Number
2050012-2011-02674
Date Received
June 26, 2011
Date of Event
April 13, 2011
Report Date
May 27, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SAMPLE AND QC INFORMATION WERE NOT PROVIDED. THE CUSTOMER STATED ONLY THAT THE QC HAD DROPPED 4-5 MMOL/L LOWER. A 4-5 MMOL/L CHANGE IN CL RESULTS IS WITHIN THE PRECISION OF THE ASSAY. ON (B)(4) 2011, A BCI FIELD SERVICE ENGINEER (FSE) EVALUATED THE INSTRUMENT AND FOUND EVIDENCE OF IMPROPER SAMPLE HANDLING AND INSUFFICIENT SPIN TIMES. THE SAMPLE INSPECTED ON THE DAY OF THE EVENT HAD CLOTS AND EVIDENCE OF GEL BEING ASPIRATED INTO THE PROBE. FSE INSTRUCTED THE CUSTOMER ON PROPER HANDLING AND THE CAUSES FOR THE EVENT. FSE REPLACED THE ELECTROLYTE INJECTION CUP (EIC) VALVES AND VERIFIED PERFORMANCE.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) IN REGARDS TO AN ERRONEOUS LOW CHLORIDE (CL) RESULTS GENERATED BY THE UNICEL DXC 800 SYNCHRON CHEMISTRY ANALYZER. THE RESULTS WERE REPORTED OUT OF THE LABORATORY. WHEN QC SHIFTED LOW, SAMPLES WERE REPEATED AND RESULTS AMENDED. THE CUSTOMER DID NOT PROVIDE INFORMATION ON THE NUMBER OF PATIENT AFFECTED OR PATIENT RESULTS. UNKNOWN IF TREATMENT WAS INITIATED OR WITHHELD BASED ON THE INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 800 SYNCHRON ® CHEMISTRY ANALYZER CLINICAL CHEMISTRY ANALYZER JJE BECKMAN COULTER INC. DXC 800 N/A

Patients

Seq Age Sex Outcome Treatment
1