FDA Adverse Event
Malfunction
Summary report: N
UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER
MDR report key: 2140234
·
Received June 26, 2011
Report
- Report Number
- 2050012-2011-02640
- Event Type
- Malfunction
- Date Received
- June 26, 2011
- Date of Event
- May 26, 2011
- Report Date
- May 26, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE CUSTOMER STATED RED LINE TUBING HAD POPPED OFF. ON(B)(6) 2011 A BCI FIELD SERVICE ENGINEER (FSE) FOUND THE REMAINS OF A LARGE LEAK ON THE INSIDE OF THE INSTRUMENT AND DETERMINED THAT THE WASTE TUBING FROM THE ACTIVE WASH TOWER WAS UNSCREWED FROM THE TOWER. FSE SCREWED THE TUBING INTO WASH TOWER AND ENSURED ALL OTHER TUBINGS WERE SECURE. FSE CLEANED INSIDE OF THE CTA AND PRIMED THE INSTRUMENT 20 TIMES WITH NO LEAKS. THE INSTRUMENT WAS OPERATIONAL.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) HOTLINE TO REPORT THE CLOSE TUBE ACCESSION (CTA) WAS FLOODED ON THE INSIDE OF THE INSTRUMENT AND ONTO THE FRONT OF THE INSTRUMENT. NO INJURY OR FUMES WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | UNICEL® DXC 600I SYNCHRON ® CHEMISTRY ANALYZER | CLINICAL CHEMISTRY ANALYZER | JJE | BECKMAN COULTER INC. | DXC 600I | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |