FDA Adverse Event Injury Summary report: N

ICL (IMPLANTABLE COLLAMER LENS)

MDR report key: 2140194 · Received June 25, 2011

Report

Report Number
2023826-2011-00565
Event Type
Injury
Date Received
June 25, 2011
Date of Event
January 19, 2011
Report Date
June 3, 2011
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P030016
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4) - SURGICAL PROCEDURE, SECONDARY SURGERY, INTRAOCULAR PRESSURE RISE, ENLARGEMENT OF PI, EXPLANTED, LENS, VAULTING, EXCESSIVE. (B)(4).

Additional Manufacturer Narrative · 1

LENS WORK ORDER SEARCH, MEDICAL REVIEW. A LENS WORK ORDER SEARCH WAS PERFORMED AND ONE SIMILAR COMPLAINT WAS FOUND WITHIN THE WORK ORDER. VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND NO VISIBLE DAMAGE TO THE LENS. THE LENS WAS RETURNED DRY. MEDICAL REVIEW - ACCORDING TO USE FMEA (FAILURE MODES AND EFFECT ANALYSIS) IT HAS BEEN DETERMINED THAT EXCESSIVE VAULTING IS A CONSEQUENCE OF A WRONG LENS USE FAILURE MODE (I.E. IMPROPER WHITE TO WHITE MEASUREMENT, VARIABILITY OF THE WHITE TO WHITE MEASUREMENTS BASED UPON DIFFERENT TECHNIQUES UTILIZED, IMPROPER SULCUS TO SULCUS MEASUREMENT (IF UBM USED), AND PATIENT CONDITION; POOR CORRELATION OF WHITE TO WHITE MEASUREMENT AND LENGTH OF CILIARY SULCUS IN AN INDIVIDUAL CASE; IRREGULAR CILIARY SULCUS OR CILIARY SULCUS CYST). SEVERAL CONDITIONS MAY ARISE FROM THIS EVENT (I.E. PUPILLARY BLOCK, MALIGNANT GLAUCOMA, INCREASED IOP ETC.). TO PREVENT ANY OF THESE COMPLICATIONS FROM OCCURRING, THE MANUFACTURER RECOMMENDS THAT THE LENS BE EXPLANTED AND/OR EXCHANGED WITH THE RIGHT SIZE ONCE THE SURGEON DETERMINES THAT THIS CONDITION MAY AFFECT THE OUTCOME OF THE PATIENT'S VISION. BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH, MEDICAL REVIEW AND THE EVALUATION OF THE RETURNED PRODUCT, THE ROOT CAUSE OF EXCESSIVE VAULTING HAS BEEN DETERMINED TO BE RELATED TO THE INACCURACY OF THE WHITE TO WHITE MEASUREMENT OR A MISMATCH BETWEEN WHITE TO WHITE AND THE SULCUS TO SULCUS DIAMETER. (B)(4).

Description of Event or Problem · 1

THE REPORTER INDICATED THE SURGEON IMPLANTED A 12.5MM ICM125V4 IMPLANTABLE COLLAMER LENS IN THE PATIENT'S RIGHT EYE (OD) ON (B)(6) 2011. THE LENS WAS EXPLANTED ON (B)(6) 2011 DUE TO EXCESSIVE VAULTING. SUBSEQUENTLY, THE PATIENT EXPERIENCED ELEVATED IOP AND THE PI WAS ENLARGED. THE ICL WAS EXCHANGED FOR A VICM 12.6 LENS WHICH RESOLVED THE PROBLEM. AT THE LAST POST-OP VISIT THE PATIENT'S BCVA WAS 20/20.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 ICL (IMPLANTABLE COLLAMER LENS) INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY ICM125V4 NA

Patients

Seq Age Sex Outcome Treatment
1 29 YR Required Intervention FOAM TIP PLUNGER: MODEL AND LOT NUMBER UNKNOWN| CARTRIDGE: MODEL AND LOT NUMBER UNKNOWN| INJECTOR: MODEL AND LOT NUMBER UNKNOWN