FDA Adverse Event Malfunction Summary report: N

UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM

MDR report key: 2140117 · Received June 25, 2011

Report

Report Number
1061932-2011-00763
Event Type
Malfunction
Date Received
June 25, 2011
Date of Event
April 20, 2011
Report Date
May 25, 2011
Manufacturer
BECKMAN COULTER, INC.
Product Code
GKZ
PMA / PMN Number
K081930
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
HR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE SPECIMENS WERE COLLECTED IN 3 ML EDTA VACUTAINER TUBES, STORED AT ROOM TEMPERATURE, AND PROCESSED WITH 2-4 HRS OF COLLECTION. RETIC CONTROLS (THREE LEVELS) WERE RUN AND WERE WITHIN QC SPECIFICATIONS FOR EACH DAY. RAW DATA FILES PROVIDED FOR THIS COMPLAINT INVESTIGATION DID NOT CONTAIN THE RETIC DATA (I.E. RETIC CYCLE) REQUIRED FOR ANALYSIS. SEVERAL ATTEMPTS WERE MADE TO OBTAIN THE ADDITIONAL INFORMATION NECESSARY FOR THE INVESTIGATION (I.E. RAW DATA, APPROPRIATE PRINTOUTS), BUT THE INFORMATION WAS NOT PROVIDED. SERVICE INFORMATION WAS REQUESTED BUT NOT PROVIDED. A DEFINITIVE ROOT CAUSE IS UNKNOWN FOR THIS EVENT. INSUFFICIENT INFORMATION WAS PROVIDED FOR THE INVESTIGATION.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) STATING THAT ONE (1) RETIC PATIENT SAMPLE RESULT OBTAINED ON THE UNICEL DXH 800 COULTER CELLULAR ANALYSIS SYSTEM AND ALTERNATE INSTRUMENTS WERE NOT COMPARABLE. PER THE CUSTOMER, ERRONEOUS RESULTS WERE REPORTED OUT OF THE LABORATORY. THE INSTRUMENT PRINTOUT CORRESPONDING TO THE PATIENT SAMPLE WAS NOT PROVIDED. AS SUCH, FLAGS, MESSAGES, AND SCATTER PATTERNS COULD NOT BE EVALUATED FOR THE SAMPLE. THE PATIENT SAMPLE HAD A HIGHER RETIC VALUE WHEN RUN ON THE DXH800 IN COMPARISON TO THE ALTERNATE INSTRUMENT. THE RESULTS PROVIDED BY THE CUSTOMER ARE PRESENTED IN ATTACHMENT PORTION OF THIS REPORT. NO REPORTS OF DEATH, INJURY, OR CHANGE TO PATIENT TREATMENT HAVE BEEN REPORTED IN CONNECTION WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXH 800 COULTER® CELLULAR ANALYSIS SYSTEM HEMATOLOGY ANALYZER GKZ BECKMAN COULTER, INC. DXH 800 N/A

Patients

Seq Age Sex Outcome Treatment
1