FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 2140047 · Received June 25, 2011

Report

Report Number
1423500-2011-08366
Event Type
Malfunction
Date Received
June 25, 2011
Date of Event
June 1, 2011
Report Date
June 1, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K923065
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ACTUAL SAMPLE WAS RECEIVED AND EVALUATED. SET WAS PLACED ON MACHINE FOR PRIMING AND RUN WITH NO ALARMS NOTED. NO MANUFACTURING ABNORMALITIES WERE NOTED DURING VISUAL INSPECTION (CAVITY 1L). THE COMPLAINT COULD NOT BE CONFIRMED IN THE LAB. THE BATCH REVIEW WAS PERFORMED ON POTENTIALLY ASSOCIATED LOT (H11D04084) WITH NO ISSUES NOTED. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).THE SAMPLE HAS NOT YET BEEN RECEIVED BUT AN EVALUATION IS EXPECTED. A FOLLOW-UP MDR WILL BE SUBMITTED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL INFORMATION IS RECEIVED.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BAXTER'S TECHNICAL SERVICE CENTER REGARDING A SYSTEM ERROR (SE) 2240 (AIR IN LINE) ALARM THAT APPEARED ON THE HOMECHOICE (HC) DISPLAY DURING DWELL 3 OF 4. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) EXPLAINED THE ALARM TO THE HOME PATIENT (HP) AND ASSISTED TO CYCLE POWER THE HC MACHINE. THE TSR ADVISED THE HP THAT THERAPY HAS ENDED AND WOULD NEED TO START OVER WITH NEW SUPPLIES OR COMPLETE THERAPY WITH MANUAL SUPPLIES. THE HP WOULD USE MANUAL SUPPLIES. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AT THE TIME OF THE INITIAL REPORT. DURING FOLLOW UP THE HOME PATIENT (H) STATED THAT AFTER STARTING OVER WITH NEW SUPPLIES NO FURTHER ISSUES WERE FOUND. THE HP STATED THIS WAS AN ISOLATED EVENT. THE HP STATED THEY DID NOT DEVELOP ANY ADVERSE REACTIONS OR NEED ANY MEDICAL INTERVENTION.