FDA Adverse Event Malfunction Summary report: N

VISIAN ICL (IMPLANTABLE COLLAMER LENS)

MDR report key: 2140036 · Received June 25, 2011

Report

Report Number
2023826-2011-00564
Event Type
Malfunction
Date Received
June 25, 2011
Date of Event
April 18, 2011
Report Date
May 31, 2011
Manufacturer
STAAR SURGICAL COMPANY
Product Code
HQL
PMA / PMN Number
P030016
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SP
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

METHOD: LENS WORK ORDER SEARCH. RESULTS: A LENS WORK ORDER SEARCH WAS PERFORMED AND NO SIMILAR COMPLAINTS WERE FOUND WITHIN THE WORK ORDER. VISUAL INSPECTION OF THE RETURNED PRODUCT FOUND A PIECE OF ONE HAPTIC BROKEN. THE LENS WAS RETURNED DRY. CONCLUSIONS: NO CONCLUSION CAN BE DRAWN. BASED ON THE COMPLAINT HISTORY, WORK ORDER SEARCH AND THE EVALUATION OF THE RETURNED PRODUCT, A SPECIFIC ROOT CAUSE OF THE EVENT COULD NOT BE DETERMINED. (B)(4).

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

THE REPORTER INDICATED THE SURGEON INSERTED A 13.2MM VICM13.2 IMPLANTABLE COLLAMER LENS IN THE PATIENT'S RIGHT EYE (OD) AND THE LENS TORE. THE LENS WAS REMOVED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VISIAN ICL (IMPLANTABLE COLLAMER LENS) INTRAOCULAR LENS HQL STAAR SURGICAL COMPANY VICM 13.2 NA

Patients

Seq Age Sex Outcome Treatment
1 26 YR FOAM TIP PLUNGER MODEL AND LOT NUMBER UNK| INJECTOR MODEL AND LOT NUMBER UNK| CARTRIDGE MODEL AND LOT NUMBER UNK