SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Report
- Report Number
- 1423500-2011-08362
- Event Type
- Malfunction
- Date Received
- June 25, 2011
- Date of Event
- June 1, 2011
- Report Date
- June 1, 2011
- Manufacturer
- BAXTER HEALTHCARE
- Product Code
- FKX
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
(B)(4). THIS COMPLAINT FOR THE SYSTEM ERROR 2240 (AIR IN LINE) WAS NOT CONFIRMED DUE TO A LACK OF SAMPLE. THE ROOT CAUSE OF THE COMPLAINT WAS NOT DETERMINED. THE LOT NUMBER IS UNKNOWN; THEREFORE A BATCH REVIEW CANNOT BE PERFORMED. SIMILAR REPORTS HAVE BEEN RECEIVED BY BAXTER FOR THE REPORTED PROBLEM. THE ROOT CAUSE INVESTIGATION IS IN PROGRESS THROUGH RENQ-CAPA-(B)(4).
(B)(4). THE SAMPLE WAS DISCARDED BY THE CUSTOMER. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF BAXTER'S INVESTIGATION, OR IF ANY ADDITIONAL INFORMATION IS RECEIVED. A 510(K) NUMBER WILL NOT BE PROVIDED IN THE EMDR, BECAUSE THE PRODUCT IS UNKNOWN AT THIS TIME.
THIS IS AN INTERNATIONAL REPORT WHERE THE HOMECHOICE MACHINE ALARMED A SYSTEM ERROR (SE) 2240 DURING DWELL 9/10. THE PATIENT WAS CONNECTED TO THE MACHINE. THE PATIENT DID NOT DISCONNECT ANY TIME PRIOR TO THE ALARM. ALL BAGS WERE PROPERLY SPIKED AND CONNECTED. THERE WERE NO OPEN CLAMPS ON UNUSED SUPPLY LINES. THERE WAS NOTHING UNUSUAL NOTED ABOUT THE SUPPLIES, I.E. LEAKS. PROPER PROCEDURES PER THE USER MANUAL WERE REVIEWED WITH THE CALLER. TECHNICAL SERVICE REPRESENTATIVE (TSR) ADVISED HOME PATIENT THAT AIR HAD ENTERED SETUP. TSR HAD HP RECYCLE POWER. TSR ADVISED HP THERAPY HAD ENDED AND WILL NEED TO START OVER WITH NEW SUPPLIES OR DO MANUAL EXCHANGE. HP WILL DO MANUAL EXCHANGE. NO CLINICAL CONSEQUENCES FOR THE PATIENT WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | HOMECHOICE |