FDA Adverse Event Summary report: N

PK7300 AUTOMATED MICROPLATE SYSTEM

MDR report key: 2140010 · Received June 25, 2011

Report

Report Number
2050012-2010-02639
Date Received
June 25, 2011
Date of Event
May 25, 2011
Report Date
May 25, 2011
Manufacturer
BECKMAN COULTER MISHIMA K.K.
Product Code
KSZ
PMA / PMN Number
BK060024
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FSE REPLACED PCB ((B)(4)) AND CALIBRATED ALL PRESSURES. FSE VERIFIED REPAIRS PER ESTABLISHED PROCEDURES. THE SYSTEM OPERATIONAL. ACCEPTABLE RESULTS OBTAINED FROM CUSTOMER POST-SERVICE VERIFICATION TEST RUN. THE ROOT CAUSE IS UNKNOWN.

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) FIELD SERVICE ENGINEER (FSE) REPORTED A BURNING SMELL WITHOUT SMOKE ON THE PK7300 AUTOMATED MICROPLATE SYSTEM. THE SMELL WAS ASSOCIATED WITH A SHORTED PRINTED CIRCUIT BOARD (PCB PART NUMBER MU7161). NO REPORTS OF DEATH OR INJURY ASSOCIATED WITH THIS EVENT. THIS EVENT DID NOT AFFECT PATIENTS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 PK7300 AUTOMATED MICROPLATE SYSTEM AUTOMATED MICROPLATE SYSTEM KSZ BECKMAN COULTER MISHIMA K.K. PK7312-02 W/OSV N/A

Patients

Seq Age Sex Outcome Treatment
1