PK7300 AUTOMATED MICROPLATE SYSTEM
MDR report key: 2140010
·
Received June 25, 2011
Report
- Report Number
- 2050012-2010-02639
- Date Received
- June 25, 2011
- Date of Event
- May 25, 2011
- Report Date
- May 25, 2011
- Manufacturer
- BECKMAN COULTER MISHIMA K.K.
- Product Code
- KSZ
- PMA / PMN Number
- BK060024
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
FSE REPLACED PCB ((B)(4)) AND CALIBRATED ALL PRESSURES. FSE VERIFIED REPAIRS PER ESTABLISHED PROCEDURES. THE SYSTEM OPERATIONAL. ACCEPTABLE RESULTS OBTAINED FROM CUSTOMER POST-SERVICE VERIFICATION TEST RUN. THE ROOT CAUSE IS UNKNOWN.
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BCI) FIELD SERVICE ENGINEER (FSE) REPORTED A BURNING SMELL WITHOUT SMOKE ON THE PK7300 AUTOMATED MICROPLATE SYSTEM. THE SMELL WAS ASSOCIATED WITH A SHORTED PRINTED CIRCUIT BOARD (PCB PART NUMBER MU7161). NO REPORTS OF DEATH OR INJURY ASSOCIATED WITH THIS EVENT. THIS EVENT DID NOT AFFECT PATIENTS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PK7300 AUTOMATED MICROPLATE SYSTEM | AUTOMATED MICROPLATE SYSTEM | KSZ | BECKMAN COULTER MISHIMA K.K. | PK7312-02 W/OSV | N/A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |