PEN NDL 32G 4MM 100BX 1200 USA
Report
- Report Number
- 3023359743-2025-00119
- Event Type
- Malfunction
- Date Received
- February 18, 2025
- Report Date
- March 13, 2025
- Manufacturer
- EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND
- Product Code
- FMI
- UDI-DI
- 00382903201228
- PMA / PMN Number
- K213478
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UT, US
- Reporter Occupation
- OTHER
- Health Professional
- N
Narratives
ADDITIONAL INFORMATION WAS ADDED TO: B4, G6, H2, H3, H11. CORRECTION TO: H6 (TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, AND INVESTIGATION CONCLUSIONS). INVESTIGATION SUMMARY: SAMPLES WERE RECEIVED AND AN INVESTIGATION WAS PERFORMED. THIS IS THE 1ST COMPLAINT FOR THE REPORTED LOT NUMBER. A REVIEW OF THE MANUFACTURING RECORDS WAS PERFORMED AND NO NON-CONFORMANCES WERE RAISED IN ASSOCIATION WITH THIS TYPE OF EVENT FOR THIS LOT. EMBECTA WAS NOT ABLE TO DUPLICATE OR CONFIRM THE INDICATED ISSUE AND BASED ON TREND ANALYSIS NO FURTHER ACTION IS REQUIRED AT THIS TIME. COMPLAINTS RECEIVED FOR THIS DEVICE AND REPORTED CONDITION WILL CONTINUE TO BE TRACKED AND TRENDED. BASED ON THE ABOVE, NO ADDITIONAL INVESTIGATION AND CORRECTIVE/PREVENTATIVE ACTION (CAPA) OR SITUATIONAL ANALYSIS (SA) IS REQUIRED.
CONSUMER REPORTED FOUND THE FIRST FEW PEN NEEDLES FROM THIS BOX TO CLOG DURING THE FLOW CHECK. TOOK PEN AND PEN NEEDLES TO PHARMACIST HE ALSO COULD NOT GET INSULIN OUT. PHARMACY PROVIDED A SAMPLE PACK OF PEN NEEDLES TO USE. DISCARDED SAMPLE OF OUR PEN NEEDLES. DC LOT # 4100049. CATALOG# 320122. DATE OF EVENT UNKNOWN SOMETIME LAST WEEK. SAMPLE STATUS DISCARD ONES WITH ISSUES. SENDING MAILING KIT FOR UNUSED SAMPLES FROM THIS BOX.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1167358 | PEN NDL 32G 4MM 100BX 1200 USA | NEEDLE, HYPODERMIC, SINGLE LUMEN | FMI | EMBECTA PENEL LIMITED - DUN LAOGHAIRE, IRELAND | 320122 | 4100049 | 00382903201228 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Prefer Not To Disclose |