FDA Adverse Event Malfunction Summary report: N

MIDAS REX LEGEND

MDR report key: 21392211 · Received February 18, 2025

Report

Report Number
1625507-2025-00317
Event Type
Malfunction
Date Received
February 18, 2025
Date of Event
May 15, 2023
Report Date
February 17, 2025
Manufacturer
MDT POWERED SURGICAL SOLUTIONS
Product Code
HBB
UDI-DI
00613994418852
PMA / PMN Number
K020069
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

H3: PRODUCT ANALYSIS: EVALUATION COULD NOT REPRODUCE THE REPORTED MALFUNCTION AS THE TEMPERATURE WAS WITHIN SPECIFICATION AT 96.4°F. THIS REGULATORY REPORT IS BEING SUBMITTED DUE TO A RETROSPECTIVE REVIEW THROUGH CAPA 624392. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

REPAIR REQUEST WAS INITIATED FOR THE DEVICE WITH THE REPORT OF OVERHEATING AND NOISY. IT WAS REPORTED THAT THERE WAS NO PATIENT INVO LVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1490512 MIDAS REX LEGEND MOTOR, DRILL, PNEUMATIC HBB MDT POWERED SURGICAL SOLUTIONS AA15 00613994418852

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown