FDA Adverse Event
Malfunction
Summary report: N
PASCO MICRO-ORGANISM SUSCEPTIBILITY SYSTEM
MDR report key: 21385
·
Received April 21, 1995
Report
- Report Number
- MW1005699
- Event Type
- Malfunction
- Date Received
- April 21, 1995
- Date of Event
- February 2, 1995
- Report Date
- February 2, 1995
- Manufacturer
- DIFCO
- Product Code
- LKA
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- WA, US
- Reporter Occupation
- PHARMACIST
Narratives
Description of Event or Problem · 1
THE INTERPRETATIONS WHICH APPEAR ON THE ATTACHED LIS REPORT, THEREFORE ARE NOT PROVIDED BY THE PASCO SOFTWARE. CONCLUSION IS THAT THE ERRONEOUS INTERPRETATIONS WERE LOADED INTO CO'S SYSTEM BY THE MICROBIOLOGY LAB PERSONNEL OF ST. JOSEPH HOSP. CO RECOMMENDED THAT ALL RESULTS AND ASSOCIATED INTERPRETATIONS AND MESSAGES BE CAREFULLY REVIEWED BEFORE CHANGES ARE MADE OR REPORTS ARE RELEASED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | PASCO MICRO-ORGANISM SUSCEPTIBILITY SYSTEM | SOFTWARE | LKA | DIFCO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | * |