TPRLC XR T1 PPS 17X154MM
Report
- Report Number
- 0001825034-2025-00368
- Event Type
- Malfunction
- Date Received
- February 14, 2025
- Date of Event
- January 22, 2025
- Report Date
- June 10, 2025
- Manufacturer
- ZIMMER BIOMET, INC.
- Product Code
- KWA
- PMA / PMN Number
- K110400
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- 003
Narratives
(B)(4). D10: 51-106170 ITEM NAME TPRLC 133 MP TYPE1 PPS SO 17.0 LOT#: 6047724. 51-107160 ITEM NAME: TPRLC 133 MP TYPE1 PPS HO 16.0 LOT#: 7118631. 51-145170 ITEM NAME: TPRLC XR MP T1 PPS 17X119MM LOT#: 3972078. 51-145170 ITEM NAME: TPRLC XR MP T1 PPS 17X119MM LOT#: 7124271. 51-104150 ITEM NAME: TPRLC 133 T1 PPS HO 15X150MM LOT#: 7178799. 51-107170 ITEM NAME: TPRLC 133 MP TYPE1 PPS HO 17.0 LOT#:7257528. G2: FOREIGN: JAPAN. CUSTOMER HAS INDICATED THAT THE PRODUCT IS IN PROCESS OF BEING RETURNED TO ZIMMER BIOMET FOR INVESTIGATION. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A FOLLOW-UP MDR WILL BE SUBMITTED.
THIS FOLLOW-UP REPORT IS BEING SUBMITTED TO RELAY ADDITIONAL INFORMATION. THE FOLLOWING SECTIONS WERE UPDATED: B4;D9;G3;H2;H3;H6. PRODUCT WAS RETURNED AND EVALUATED. VISUAL EVALUATION OF THE RETURNED PRODUCT/PROVIDED PHOTOS IDENTIFIED DAMAGE TO THE STERILE PACKAGING (BLISTER). STERILITY HAD NOT BEEN COMPROMISED. REVIEW OF THE DEVICE HISTORY RECORDS IDENTIFIED NO DEVIATIONS OR ANOMALIES DURING MANUFACTURING RELATED TO THE REPORTED EVENT. REPORTED EVENT IS NOT RELATED TO A COMBINATION OF PRODUCTS; THEREFORE, A COMPATIBILITY REVIEW IS NOT APPLICABLE. THE REPORTED EVENT DID NOT OCCUR IN AN OPERATING ROOM OR AS PART OF A MEDICAL PROCEDURE; MEDICAL RECORDS ARE NOT AVAILABLE FOR REVIEW. THE CONDITION OF THE DEVICE WHEN IT LEFT ZIMMER BIOMET IS CONSIDERED CONFORMING TO SPECIFICATION. THE ROOT CAUSE OF THE REPORTED EVENT CAN BE ATTRIBUTED TO TRANSIT DAMAGE AND A PACKAGING DESIGN ISSUE. THE REPORTED EVENT HAS BEEN CONFIRMED BY EVALUATION OF THE RETURNED PRODUCT AND PROVIDED PHOTOS. IF ANY FURTHER INFORMATION IS FOUND WHICH WOULD CHANGE OR ALTER ANY CONCLUSIONS OR INFORMATION, A SUPPLEMENTAL WILL BE FILED ACCORDINGLY. ZIMMER BIOMET WILL CONTINUE TO MONITOR FOR TRENDS.
IT WAS REPORTED THAT WHEN INSPECTING THE PRODUCT, THE STERILE PACKAGING WAS FOUND DAMAGED. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1488924 | TPRLC XR T1 PPS 17X154MM | PROSTHESIS, HIP | KWA | ZIMMER BIOMET, INC. | 7206106 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |