FDA Adverse Event Malfunction Summary report: N

E-KATHETER II SET TSUI

MDR report key: 21383088 · Received February 14, 2025

Report

Report Number
9611612-2025-00004
Event Type
Malfunction
Date Received
February 14, 2025
Date of Event
January 5, 2025
Report Date
April 25, 2025
Manufacturer
PAJUNK GMBH MEDIZINTECHNOLOGIE
Product Code
BSO
UDI-DI
04048223121931
PMA / PMN Number
K152952
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
RISK MANAGER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

IRN#(B)(4). COMPLAINT TOOK PLACE IN USA. UNDER ITEM H6, CODES B, C AND D, PLACEHOLDERS HAVE BEEN INSERTED THAT HAVE NOT BEEN VERIFIED, AS THE INTIAL REPORT CAN ONLY BE PACKED IN THIS FORM IF THESE ARE FILLED IN, ALTHOUGH NO ANALYSIS HAS YET TAKEN PLACE. THESE B, C AND D CODES CAN ONLY BE TRANSMITTED CORRECTLY IN THE SUBSEQUENT MESSAGE AS THE ANALYSIS IS STILL BEING PROCESSED. YOU WILL RECEIVE THE COMPLETE REPORT WITH ALL TEST RESULTS IN THE SUBSEQUENT MESSAGE.

Additional Manufacturer Narrative · 0

IRN # (B)(4). COMPLAINT TOOK PLACE IN USA. BASED ON RISK ASSESSMENT AND CLINICAL ASSESSMENT THIS FILE IS CONSIDERED AS CLOSED. IN CASE NEW INFORMATION BECOMES AVAILABLE A FOLLOW UP REPORT WILL BE SENT TO THE AGENCY.

Description of Event or Problem · 0

IRN# (B)(4). INCIDENT OCCURRED IN USA: WITH NEEDLE INSERTED, UPON INJECTING MEDICATION, THE "HUB" SPRUNG LEAK. THERE WAS A "FOUNTAIN" OF MEDICATION SQUIRTING FROM THE HUB WHILE UNDER PRESSURE. TRIED ANOTHER NEEDLE AND THE SAME EVENT OCCURED. TRIED A DIFFERENT PKG., DIFFERENT LOT#- ALL OK.

Description of Event or Problem · 0

IRN # (B)(4). INCIDENT OCCURRED IN USA: WITH NEEDLE INSERTED, UPON INJECTING MEDICATION, THE "HUB" SPRUNG LEAK. THERE WAS A "FOUNTAIN" OF MEDICATION SQUIRTING FROM THE HUB WHILE UNDER PRESSURE. TRIED ANOTHER NEEDLE AND THE SAME EVENT OCCURED. TRIED A DIFFERENT PKG., DIFFERENT LOT#- ALL OK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1415759 E-KATHETER II SET TSUI E-KATHETER II SET TSUI_CATHETER, CONDUCTION, ANESTHETIC BSO PAJUNK GMBH MEDIZINTECHNOLOGIE AA610027 1562 04048223121931

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Other