BD BBL¿ TB DECOLORIZER
Report
- Report Number
- 1119779-2025-00072
- Event Type
- Malfunction
- Date Received
- February 13, 2025
- Date of Event
- January 3, 2025
- Report Date
- February 28, 2025
- Manufacturer
- BECTON DICKINSON & CO. (SPARKS)
- Product Code
- LED
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS ANTICIPATED BUT IS NOT COMPLETE. UPON COMPLETION, A SUPPLEMENTAL REPORT WILL BE FILED.
INVESTIGATION SUMMARY - THIS MEMO IS TO SUMMARIZE FINDINGS ON THE COMPLAINT RELATED TO BOTTLE TB DECOLORIZER 250 ML, CATALOG NUMBER 212517, BATCH NUMBER 4100005, AND COMPLAINT NUMBER (B)(4) FOR PERFORMANCE. COMPONENTS ARE MIXED AND DISPENSED INTO CONTAINERS. FOLLOWING QC RELEASE, AND BASED UPON INVENTORY DEMAND, FINAL PACKAGING OPERATIONS OCCUR FOLLOWED BY TRANSPORT TO THE DISTRIBUTION CENTER. THE COMPLAINT INVESTIGATION INCLUDED A REVIEW OF THE BATCH HISTORY RECORD FOR TB DECOLORIZER. THE BATCH HISTORY RECORD REVIEWS INDICATED NO DISCREPANCIES. ALL RELEASE TESTING WAS SATISFACTORY. COMPLAINT TRENDS WERE REVIEWED FOR A PERIOD COVERING 12 MONTHS. DURING THAT TIME THERE HAVE BEEN NO CONFIRMED COMPLAINTS FOR THE DEFECT IN QUESTION. TWO PHOTOS WERE ATTACHED IN LIEU OF RETURN. ALL PHOTOS DEPICT A DIFFERENT MICROSCOPE VIEW OF ACID-FAST STAIN BACTERIA. BD BBL¿ TB STAIN KITS ARE USED TO MANUALLY STAIN SMEARS PREPARED FROM SPECIMENS AND CULTURES SUSPECTED OF CONTAINING MYCOBACTERIA, ESPECIALLY MYCOBACTERIUM TUBERCULOSIS, FOR QUALITATIVE EARLY PRESUMPTIVE DIAGNOSIS OF MYCOBACTERIAL INFECTION AND CHARACTERIZATION OF BACTERIAL ISOLATES. SLIDES OF POSITIVE (M. TUBERCULOSIS ATCC 25177) AND NEGATIVE (B. SUBTILIS ATCC 6533) CONTROLS WERE EVALUATED AND GAVE SATISFACTORY STAINING RESULTS AT RELEASE OF THE PRODUCT. FROM CUSTOMER'S RESPONSE: ¿AFB QC SLIDES WERE USED AND PRODUCED EXPECTED RESULTS. POSITIVE CONTROL SLIDE HAD PINK/RED ACID-FAST BACILLI. NEGATIVE CONTROL SLIDE HAD GREENISH-BLUE BACKGROUND¿. THE ORGANISM DESCRIBED IN THE COMPLAINT REPORT IS NOCARDIA. BD DOES NOT HAVE ESTABLISHED RELEASE REQUIREMENTS OR PERFORMANCE DATA FOR STAINING NOCARDIA. BASED ON THE LIMITATIONS OF THE PROCEDURE AND RELEASE SPECIFICATIONS FOR THE PRODUCT, WE ARE UNABLE TO FURTHER REPLICATE THE CUSTOMER¿S APPLICATION WITH THIS PRODUCT. THIS COMPLAINT CANNOT BE CONFIRMED. BD WILL CONTINUE TO TREND ON PERFORMANCE ISSUES.
IT WAS REPORTED WHILE USING BD BBL¿ TB DECOLORIZER, ONE PATIENT SAMPLE CONTAINED BRANCHING ORGANISMS THAT WAS STAINING KINYOUN POSITIVE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
IT WAS REPORTED WHILE USING BD BBL¿ TB DECOLORIZER, ONE PATIENT SAMPLE CONTAINED BRANCHING ORGANISMS THAT WAS STAINING KINYOUN POSITIVE. THERE WAS NO HEALTH IMPACT OR CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1489885 | BD BBL¿ TB DECOLORIZER | STAINS, CHEMICAL SOLUTION | LED | BECTON DICKINSON & CO. (SPARKS) | 4100005 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |