FDA Adverse Event Injury Summary report: N

MENTOR MEMORYGEL BREAST IMPLANT

MDR report key: 21381881 · Received February 13, 2025

Report

Report Number
1645337-2025-01561
Event Type
Injury
Date Received
February 13, 2025
Date of Event
March 8, 2017
Manufacturer
MENTOR TEXAS
Product Code
FTR
UDI-DI
00081317000297
PMA / PMN Number
P030053
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NC, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

ON FEBRUARY 14, 2025, MENTOR RECEIVED ADDITIONAL INFORMATION INDICATING THAT THE PATIENT'S IMPLANTS WERE REPLACED WITH THE FOLLOWING DEVICES: (RIGHT) 375CC MENTOR MEMORYGEL BREAST IMPLANT, CATALOG: 3503754BC, LOT: 9925030, SN: (B)(6) AND (LEFT) 375CC MENTOR MEMORYGEL BREAST IMPLANT, CATALOG: 3503754BC, LOT: 9884485, SN: (B)(6).

Additional Manufacturer Narrative · 0

SINCE THE DEVICE HAS NOT BEEN RETURNED FOR ANALYSIS, NO PRODUCT FAILURE ANALYSIS CAN BE CONDUCTED, AND NO DETERMINATION OF POSSIBLE CONTRIBUTING FACTORS CAN BE MADE. AS SUCH, THE INVESTIGATION WILL BE CLOSED. IF THE COMPLAINT DEVICE IS RECEIVED IN THE FUTURE, THE INVESTIGATION WILL BE REOPENED AND CONDUCTED AS APPROPRIATE. A MANUFACTURING RECORD EVALUATION (MRE) WAS PERFORMED, AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, THE MRE VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. REASON FOR DEVICE EXPLANT AND/OR REOPERATION: CAPSULAR CONTRACTURE MENTOR IS SUBMITTING THIS REPORT PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803. THIS REPORT MAY BE BASED ON INFORMATION WHICH MENTOR HAS NOT BEEN ABLE TO INVESTIGATE OR VERIFY PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY FDA, MENTOR, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE DEVICE, MENTOR, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. MANUFACTURER¿S REFERENCE NUMBER: (B)(4).

Additional Manufacturer Narrative · 0

ON MARCH 3, 2025, THE MENTOR FAILURE ANALYSIS LAB RECEIVED THE DEVICE FOR EVALUATION. ON MARCH 10, 2025, THE PRODUCT INVESTIGATION WAS COMPLETED. DEVICE INVESTIGATION SUMMARY: THE PRODUCT WAS RETURNED TO MENTOR FOR EVALUATION. MENTOR CONDUCTED A VISUAL INSPECTION OF THE RETURNED DEVICE. DURING THE VISUAL EVALUATION, NO APPARENT DAMAGE OR VISUAL ANOMALIES WERE OBSERVED ON THE SM MPP GEL 325CC RETURNED DEVICE. REGARDING THE REPORTED CAPSULAR CONTRACTURE CONDITION, THERE ARE NOT ENOUGH DETAILS TO DETERMINE THE FACTORS THAT MAY HAVE CAUSED THIS CONDITION. CAPSULAR CONTRACTURE IN THE PATIENT¿S BREAST MIGHT BE THE RESULT OF THE BODY¿S INDIVIDUAL PHYSIOLOGICAL RESPONSE TO THE IMPLANTATION OF A FOREIGN OBJECT IN SOFT TISSUE. FURTHER, CAPSULAR CONTRACTURE IS A KNOWN COMPLICATION ASSOCIATED WITH THESE DEVICES AND IS REFERENCED IN OUR CURRENT PRODUCT INSERT DATA SHEET. THE AMERICAN SOCIETY OF PLASTIC SURGEONS RECOMMENDS AND ENCOURAGES MEMBER SURGEONS TO ALWAYS SUBMIT BREAST IMPLANTS, CAPSULE, AND EFFUSION TO PATHOLOGY FOR EXAMINATION. THE CONTRALATERAL DEVICE WAS ALSO RECEIVED (LOT NUMBER 7328528). THE PATIENT DID NOT REPORT ANY ISSUE WITH THIS DEVICE. THEREFORE, NO FURTHER ANALYSIS OF THE CONTRALATERAL DEVICE IS REQUIRED AS PART OF MENTOR'S QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED ACCORDING TO APPROVED SPECIFICATIONS. BASED ON THE RESULTS OF THIS INVESTIGATION, IT WAS DETERMINED THAT THE PRODUCT MET THE MANUFACTURING RELEASE CRITERIA: SINCE NO MALFUNCTION WAS OBSERVED DURING THE INVESTIGATION. NO CORRECTIVE AND PREVENTIVE ACTION (CAPA) IS REQUIRED NOW.

Description of Event or Problem · 0

IT WAS REPORTED THAT A PATIENT UNDERWENT PRIMARY BREAST AUGMENTATION WITH TWO 325CC MENTOR MEMORYGEL BREAST IMPLANTS. POST-OPERATIVELY, THE PATIENT WAS DIAGNOSED, VIA PHYSICAL EXAMINATION, WITH RIGHT BREAST CAPSULAR CONTRACTURE (BAKER GRADE II-III). AS A RESULT, THE PATIENT UNDERWENT BREAST IMPLANT REMOVAL SURGERY ON (B)(6) 2025.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
982270 MENTOR MEMORYGEL BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR MENTOR TEXAS 6862958 00081317000297

Patients

Seq Age Sex Outcome Treatment
1 32 YR Female Required Intervention