INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2025-02541
- Event Type
- Injury
- Date Received
- February 13, 2025
- Date of Event
- December 18, 2024
- Report Date
- April 2, 2025
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS RUPTURE WAS RECEIVED ON FEBRUARY 27, 2025 WITH LOT NUMBER 3021914. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: RUPTURE: OBSERVED A BROKEN ASSESSED AS AN UNIDENTIFIED (TEAR) OPENING (SHELL THICKNESS WITHIN SPEC), OBSERVED A BROKEN AND OPENING ASSESSED AS SURGICAL DAMAGE AND OBSERVED MISSING SHELL ASSESSED AS INCONCLUSIVE. NO OTHER OBSERVATIONS OBSERVED, NO FURTHER ACTIONS ARE REQUIRED.
A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. DEVICE WAS EXPLANTED AND REPLACED WITH ANOTHER MANUFACTURER'S DEVICE.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. DEVICE WAS EXPLANTED AND REPLACED WITH ANOTHER MANUFACTURER'S DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1125009 | INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 3021914 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 35 YR | Female | Required Intervention |