FDA Adverse Event Injury Summary report: N

INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 21381338 · Received February 13, 2025

Report

Report Number
9617229-2025-02541
Event Type
Injury
Date Received
February 13, 2025
Date of Event
December 18, 2024
Report Date
April 2, 2025
Manufacturer
ALLERGAN (COSTA RICA)
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS RUPTURE WAS RECEIVED ON FEBRUARY 27, 2025 WITH LOT NUMBER 3021914. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINTS ARE: RUPTURE: OBSERVED A BROKEN ASSESSED AS AN UNIDENTIFIED (TEAR) OPENING (SHELL THICKNESS WITHIN SPEC), OBSERVED A BROKEN AND OPENING ASSESSED AS SURGICAL DAMAGE AND OBSERVED MISSING SHELL ASSESSED AS INCONCLUSIVE. NO OTHER OBSERVATIONS OBSERVED, NO FURTHER ACTIONS ARE REQUIRED.

Additional Manufacturer Narrative · 0

A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. DEVICE WAS EXPLANTED AND REPLACED WITH ANOTHER MANUFACTURER'S DEVICE.

Description of Event or Problem · 0

HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. DEVICE WAS EXPLANTED AND REPLACED WITH ANOTHER MANUFACTURER'S DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1125009 INSPIRA TEXTURED SILICONE GEL FILLED BREAST IMPLANT PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED FTR ALLERGAN (COSTA RICA) 3021914

Patients

Seq Age Sex Outcome Treatment
1 35 YR Female Required Intervention